Skip to main content

Select your region & language

Global

Region

Laser Doppler Vibrometer
LV-1800

Unending evolution deepens non-contact vibration measurement.

The LV-1800 Laser Doppler Vibrometer achieves more than 10 times the sensitivity compared to conventional products due to a newly designed interference optical system, and it is a non-contact vibrometer that allows for confirmation of the laser irradiation position with a built-in camera.

In recent years, small actuators driven by piezoelectric elements, as well as mobile phones, smartphones, and digital cameras equipped with MEMS (Micro Electro Mechanical Systems), have become widespread, and in the automotive sector, inverters and peripheral devices used in EV/HEV vibrate at over 20 kHz.
However, it is difficult to measure vibrations of small parts or in the ultrasonic range using contact-type acceleration pickups, which are widely used for existing vibration detection.

By using the non-contact LV-1800, it becomes possible to quickly and easily detect vibrations. Furthermore, by combining it with a waveform analysis device (FFT analyzer), it is possible not only to measure vibrations but also to accurately capture, visualize, and analyze the state of the vibrations.

Features

Integrating a high-sensitivity camera with a sensor.

You can display and confirm the target object or the area where the laser light is applied on a Windows PC without any loss of sensitivity.

Significantly improved detection sensitivity.

A new design of the interferometric optical system has been established, achieving a 10-fold (+20 dB) increase in detection sensitivity compared to conventional models. The range of detectable targets has been expanded, and the constraints on sensor installation and setup have been relaxed.

Measuring targets over a wide range with a maximum of 4 ranges.

The detectable amplitude has become 0.05 µm/s (*) to 10 m/s. It can detect a wide range of behaviors from the high-speed amplitudes of piezoelectric elements to the extremely small amplitudes emitted by thin films, MEMS, and ceramic capacitors. (*)…when the LV-0800 micro velocity range board is attached.

Easily grasp the focus position and detection status at a glance.

Equipped with a level indicator on the sensor. Settings can be made while grasping the focus position and detection status at hand.

Storage and movement are easy.

Compared to previous models, the main unit has been made 33% lighter. It also features sensor housing and cable winding, allowing for safe storage and transport. The optional storage trunk can hold not only the main unit but also accessories like a magnetic stand and lighting unit.

Powerful solution with options + measurement

By using a rich array of dedicated options, it enables the realization of various detection scenes. Additionally, the waveform analysis device and software visualize the behavior and characteristics of the target from the waveform, providing a powerful and smooth solution.

Video

Non-contact Vibration/Speed Measurement Seminar Useful for Evaluating Electric Products

(21:26)

Case Studies

ultrasonic application devices

Ultrasonic welder, bonding machine, ultrasonic cutter
Ultrasonic homogenizer, ultrasonic soldering iron
Ultrasonic microscope, ultrasonic sprayer
Fish finder, sonar device frequency, amplitude measurement
Nano-imprinting device, piezoelectric vibrating gyro
Ultrasonic motor, ultrasonic cleaner, ultrasonic scalpel
Detection of biological parts and insect behavior

IT and media-related

Transmission characteristics of optical pickups for CD, DVD, BD, etc.
Measurement of disk surface wobble
Behavior of inkjet elements
Measurement of HDD sliders & VCM, vibration analysis of flying heads, and each component
Microphones, receivers, and speaker thin film vibrations of mobile phones and smartphones
Measurement of vibration motors
Ultrasonic motors for camera AF (auto-focus)

Automobiles (including EVs and HEVs)

Inverter motor, inverter and configuration devices
Smoothing capacitors and other peripheral vibration analysis
Engine and transmission rotational tracking analysis
Shock sensor (MEMS)
Internal vibration of light bulbs
Measurement of operating displacement of brake calipers and fuel injectors
Vibration of the horn sound

Plant

Valve behavior
Long-distance vibration detection of piping, blowers, motors, etc.
Laser ultrasound

Laser Doppler Vibrometer Measurement Examples

Laser Doppler Vibrometer Measurement Examples

 

System Configurations

Illustration (LV-1800 System Configuration) LV-0930 変位解析ソフトウェア

*1 Image display is when the LV-0181 positioning built-in camera (OP) is installed.
*2 For details, please refer to the product introduction of various FFT analyzers.

Specification

detection section

    LV-1800
Demodulation method Optical Heterodyne Detection speed recovery
laser light Light source He-Ne laser λ = approximately 633 nm
injection output 1 mW less than
Laser Safety Standards Laser Class 2 Please refer to the applicable standards.
Minimum laser spot diameter approximately 20 µm or less Focal position at 100 mm φ = 1/e²
Below 3 µm (when equipped with LV-0151B) WD = approximately 38.8 mm

standard lens

variable focus type 100 mm to 10 m (∞)
distance scale 100 mm to 10 m (∞) There is an interference distance mark.
attachment size M22×0.5 depth 5.5 mm
Built-in positioning camera
LV-0181 (optional)
wearing Embedded within the sensor Factory-installed
*Can be retrofitted upon pickup
interface USB2.0 Output from the conversion section via USB mini-B connector.
imaging sensor 1/4 inch type color CMOS
number of pixels approximately 300,000 pixels
image size VGA 640×480
Frame rate 30 frames per second
Minimum imaging range 10 mm × 7.5 mm (TYP) WD = 100 mm (minimum)
2.1 mm × 1.6 mm (TYP) When the objective lens is attached to LV-0151B
imaging position Upright image (rotatable) when facing the indicator face.
Exposure automatic
white balance automatic
Gain automatic
operating environment Windows® 10/11 Display True Color greater than 24 bit
camera focus Adjustment at the objective lens Laser spot and confocal
Detection unit suspension screw hole for suspension Back part x1 M8 depth 8 mm
※Compatible with LV-0030 large magnet stand
side part x 1
side parts x2 M4 depth 5 mm
tripod setup Use of LV-0019 camera screw adapter. Option
Recovery Sensitivity Monitor Signal Level Indicator 10-segment LED array display linked with the display of the conversion section level indicator
ERROR Indicator LED display (red color)
Signal cable cable length 3 m Winding storage in the conversion section.
thickness φ = 10.5 mm
covering oil-resistant coating
minimum bending radius R = 40 mm or more
External dimensions of the detection part. Width 53 mm excluding the protruding part
Height 52.5 mm
Depth 152.5 mm
Weight of the detection unit   約750 g In the equipment state of option (LV-0181) / excluding cables

Conversion section

    LV-1800
detection speed Frequency Range 0.3 Hz to 3 MHz (fc=-3dB) *Common for all speed ranges
0.001 (m/s)/V (optional) 0.3 Hz to 200 kHz (fc=-3 dB)
Maximum detection speed 10 m/s 0-p (20 m/s p-p)
minimum speed resolution Below 0.3 µm/s (0.01 (m/s)/V) Below 0.05 µm/s (when LV-0800 is installed)
Output ±10 V (20 V p-p) When the input impedance is 100 kΩ or more.
Amplitude versus Output Voltage Polarity When approaching the sensor side + voltage
DC offset within 20 mV
output impedance 50 Ω
minimum input impedance more than 100 kΩ
Connector type BNC (C02 type)
speed range 1.0 (m/s)/V 10 m/s 0-p (20 m/s p-p)
0.1 meters per second per volt 1 m/s 0-p (2 m/s p-p)
0.01 (m/s)/V 0.1 m/s 0-p (0.2 m/s p-p)
0.001 (m/s)/V (option) 0.01 m/s 0-p (0.02 m/s p-p) ※For details, refer to the LV-0800 micro speed range board.
Over Indicator Detection speed limit +5% Over activates red LED.
Recovery Sensitivity Monitor Signal Level Indicator 20 Segment LED Array Display Linked with the detection unit level display
Monitor output 0 to 10 V
output impedance 50 Ω
minimum input impedance more than 100 kΩ
Connector type BNC (C02 type)
ERROR Indicator LED (red) lighting indication
High-pass filter (HPF) 100 Hz fc=-3dB
OFF(0.3 Hz)
low pass filter 50 kHz fc=-3dB

※ Not selectable when using the 0.001(m/s)/v range
100 kHz
1 MHz ※
OFF (3 MHz)
Positioning video output
(optional)
video output digital
Standard USB2.0
Display LED (white) lights up *When attached to LV-0181*
connector type USB mini-B type
Laser Radiation Control ON/OFF via the front panel switch The specification for starting laser radiation upon power-on (panel mode) is possible with the factory settings.
Laser ON Indicator Laser Emission LED (Green)
light-blocking shutter Contact Input Stop laser emission by opening the contact.
Connector type Receptacle: RM12BRB-2S
Plug: RM12BPE-2PH (short-circuited)
storage device Detection Unit Storage Store in the conversion section.
Cable storage Winding storage in the conversion section.
Outline Dimensions Width 410 mm excluding each protruding part
Height 120 mm
Depth 324 mm
Weight 約8.1 kg including sensor cables
Operating temperature range 0 to 40 °C
Operating Humidity Range 30 to 80 % RH However, it should not condense.
Storage Temperature Range -10 to 50 °C
input voltage AC 100~240 V
50/60 Hz
power consumption 60 VA
cooling method Natural air cooling (vibration-free cooling)

Applicable Standards

    LV-1800
Applicable Standards Laser safety IEC 60825-1:2007:2014
EN 60825-1:2007:2014/A11:2021
FDA (CDRH) 21CFR 1040.10 and 1040.11 except for conformance with IEC 60825-1 Ed.3., as described in Laser Notice No. 56.
CE marking Low Voltage Directive 2014/35/EU Standard EN 61010-1
EN 60825-1
EMC instruction 2014/30/EU Standard EN 61326-1
RoHS Directive 2011/65/EU Standard EN IEC 63000
  • Target amplitude: Polarity of output voltage

    Target amplitude: Polarity of output voltage

    <Resolution and Dynamic Range Evaluation Conditions>
    LPF: 100 kHz ON
    Maximum demodulation using corner cube
    FFT analyzer power spectrum observation
    1 kHz range 1 kHz, 2048 lines, 256 times averaging
    ※ Please refer to the graph for the frequency characteristics of each filter.

  • LV-0181 Built-in Camera Imaging Size for Positioning

    LV-0181 Built-in Camera Imaging Size for Positioning
  • The relationship between operating distance and spot diameter

    The relationship between operating distance and spot diameter

Option

(The following options are not applicable for CE marking.)

LV-0930 Analysis Software

The LV-0121A digital displacement meter is equipped with a digital displacement output. By using the LV-0930 displacement analysis software, the vibration amplitude detected by the LV-1800 + LV-0121A digital displacement meter can be output via USB 2.0 as digital displacement values (m) with a maximum sampling rate of 1 MHz and a resolution of 0.155 nm for analysis. On the LV-0930 displacement analysis software, you can perform conversions from displacement (m) to velocity (m/s) and acceleration (m/s 2), as well as moving averages, differential operations, and FFT analysis without signal degradation. Various waveform conversions and analysis results can be exported as CSV files. By importing the CSV data into spreadsheet software or the OS-2000 series time-series data analysis tools, even more advanced data processing can be performed.

  • LV-0930 Analysis Software

Features

  • Maximum 2.5 m/s, supports 1 MHz sampling.

  • "Simultaneous conversion between displacement, velocity, and acceleration is also easy."

use

  • By using the LV-0930 displacement analysis software, it is possible to analyze the dynamic behavior that occurs when the measurement target moves and stops at a sampling rate of up to 1 MHz.

LV-0181 Built-in Camera for Positioning

The built-in positioning camera LV-0181 is a highly sensitive digital camera integrated into the LV-1800 sensor. The coaxial and confocal camera, which shares the focus of the laser light and the image, outputs via USB 2.0 to display the detected area on a Windows® PC, allowing for quick confirmation of small measurement targets and laser irradiation. It is also compatible with the LV-0151B objective lens and the LV-0185 illumination unit.

  • LV-0181 Built-in Camera for Positioning

Connector type: USB2.0 Mini-B type on the device side
Imaging sensor: CMOS color sensor 1/4 inch
Pixel count: 300,000 pixels or more
Image size: VGA (640×480)
Frame rate: 30 frames/second
Capture range: 10 mm × 7.5 mm ~ (Measurement distance 100 mm ~)
Functions: Exposure/Gain/White balance (automatic)
Operating environment: Standard compatible with Windows® 10/11
Display True Color 24bit or more
Camera focus: Shared with laser light focus
Accessories: LV-0181 Camera Monitor software CD-ROM
USB cable (CF-0703) 1.5 m

(User registration is required for downloading.)

LV-0800 Micro Speed Range Board

This is a micro-speed range board for the LV-1800. By adding a range of 0.001(m/s)V (±0.01 m/s (MAX)) to the LV-1800, it enables the detection of vibrations that are difficult to measure with standard ranges, such as the amplitude of ceramic capacitors and the propagation of ultrasound in materials. With the addition of the LV-0800, it covers speed amplitudes from 0.05 μm/s to 10 m/s across four ranges.

LV-0800 Micro Speed Range Board

Speed range: 0.001 (m/s)/V (0.01 m/s 0-p (MAX))
Speed resolution: 0.05 µm/s ※ at maximum demodulation
Frequency range: 0.3 Hz to 200 kHz (fc = -3 dB)

LV-0112 Displacement Output Board / LV-0111 Acceleration Output Board

LV-0112 Displacement Output Board / LV-0111 Acceleration Output Board

Signal source: Receives LV-1800 speed signal internally
Output voltage: ±10 V (MAX)
DC offset: Within 20 mV
Maximum amplitude: 10 times of each set range (0-p)
Amplitude conversion error: Within ±5%
Amplitude output polarity: Positive voltage when approaching the sensor side

LV-0112 Displacement Output Board

LV-1800
Setting Range
Displacement Setting Range
1 Hz to 20 kHz 10 Hz to 50 kHz 1 kHz to 200 kHz
1.0(m/s)/V 100 mm/V 1 mm/V 10 micrometers per volt
0.1 (m/s) / V 10 mm/V 100 µm/V 1 micrometer per volt
0.01 (m/s) per V 1 mm/V 10 micrometers per volt 100 nm/V
0.001(m/s)/V 0.1 mm/V 1 micrometer per volt 10 nm/V

LV-0111 Acceleration Output Board

LV-1800
Setting Range
Acceleration setting range
1 Hz to 2 kHz 1 Hz to 20 kHz 100 Hz to 400 kHz
1.0(m/s)/V 10 3 (m/s 2)/V 10 5 (m/s 2)/V 10 7 (m/s 2)/V
0.1 (m/s) / V 10 2 (m/s 2)/V 10 4 (m/s 2)/V 10 6 (m/s 2)/V
0.01 (m/s) per V 10 1 (m/s 2)/V 10 3 (m/s 2)/V 10 5 (m/s 2)/V

LV-0185 Lighting Unit

This is an option for illuminating the detection target using laser light and coaxial reflection. When the objective lens is attached, it focuses the illumination on the detection area at the same working distance, enhancing the clarity of images from the built-in positioning camera LV-0181. It also makes it easier to irradiate and position lasers in recessed areas or small detection sites that are not illuminated.

LV-0185 Lighting Unit

Compatible lenses: LV-0150B (5×) / LV-0151B (10×)
LV-0152B (20×)
Illumination method: White cold light coaxial incident illumination by LED
Pixel count: 300,000 pixels or more
Cable length: 1.5 m (when using the included extension cable)
Light source: Cold light illumination by white LED
Control: Variable adjustment
Operating temperature range: 0 to 40 ℃ (no condensation)
Operating humidity range: 30 to 80 %RH (no condensation)
Input voltage: AC 100 V to 240 V 50/60 Hz
Power consumption: 3.5 VA at AC 100 V / 9.0 VA at AC 240 V

Vibration and Measurement System for Small Objects

By equipping the LV-1800 with the built-in camera for positioning LV-0181, the LV-0185 lighting unit, and various objective lenses, it enables the micro-spotting of laser light and image observation through incident illumination. This is a system that allows for laser positioning on microstructures and detection of multiple points.

LV-0381 Microscope Unit

The LV-0381 microscope unit is an optical unit for the Laser Doppler Vibrometer. It enables the detection of vibrations from MEMS, piezoelectric elements, microstructures, and ultrasonic-driven targets. Each unit allows for quick and accurate positioning of the laser light and still image recording using coaxial illumination and two digital video cameras with different magnifications (optional).

Three-Dimensional Optical Unit

The LV-3800 three-dimensional optical unit is a three-dimensional unit that enables non-contact detection of vibrations occurring in the XYZ directions using a Laser Doppler Vibrometer. It allows for the detection of vibration extension states and shear direction vibrations of ultrasonic devices and elements, which are difficult to detect using contact methods, as well as the vibrations of small objects.

  • Three-Dimensional Optical Unit

LV-0383 Three-Dimensional Microscope Unit

The LV-0383 three-dimensional microscope unit is an optical unit for the Laser Doppler Vibrometer. With three Laser Doppler Vibrometers and a microscope device, it can non-contact detect the vibrations of small objects occurring in three axial directions. Coaxial illumination and two digital video cameras with different magnifications (optional) enable quick and accurate positioning and still image recording. It allows for the evaluation of the three-dimensional behavior of MEMS (Micro Electro Mechanical Systems), microstructures, small mass targets, and ultrasonic-driven target vibrations that are difficult to detect by contact methods.

  • Objective Lens LV-0150B

    Objective Lens LV-0150B

    Magnification: 5x
    WD: 36.1 mm
    Spot diameter: ⌀4 μm or less
    Includes conversion adapter

  • Objective lens LV-0151B

    Objective lens LV-0151B

    Magnification: 10x
    WD: 38.8 mm
    Spot diameter: ⌀3 μm or less
    Includes conversion adapter

  • Objective Lens LV-0152B

    Objective Lens LV-0152B

    Magnification: 20x
    WD: 22.5 mm
    Spot diameter: ⌀2.5 μm or less
    Includes conversion adapter

  • Large Magnet Stand LV-0030

    Large Magnet Stand LV-0030

    This is a dedicated magnetic stand for sensor suspension. It allows for free laser irradiation with a cross clamp. When combined with micro stages LV-0015/LV-0016, it enables fine adjustments of the detection position.

  • Slim Type 90-Degree Reflective Mirror LV-0105

    Slim Type 90-Degree Reflective Mirror LV-0105

    Convert the tip of the 90-degree reflection mirror to a diameter of 4 mm. Rod tip diameter: 4 mm.

  • 90-degree reflection mirror LV-0301

    90-degree reflection mirror LV-0301

    By attaching it to the tip of the LV-1800 lens, you can bend the laser light by 90 degrees and rotate it 360 degrees. It is possible to detect recessed areas by inserting a mirror into gaps. Rod tip diameter: ⌀10 mm.

  • XY Axis Micro-Movement Stage LV-0015

    XY Axis Micro-Movement Stage LV-0015

    The LV-0030 can be combined with a large magnet stand for precise positioning in the XY direction. It can also be used for positioning samples when used alone.
    Stage surface: 60 × 60 mm
    Movement range: ±6.5 mm

  • Z-axis fine motion stage LV-0016

    Z-axis fine motion stage LV-0016

    The LV-0030 allows for slight vertical movement when combined with a large magnet stand. It makes focusing laser light and images, as well as positioning, easy to perform.
    Stage surface: 60 × 60 mm
    Movement amount: 0–13 mm

  • Steel Plate LV-0018A

    Steel Plate LV-0018A

    The LV-0030 large magnet stand can be used on a large tripod head or on a non-magnetic table. Additionally, by directly securing the LV-0015/0016 fine adjustment stages with screws, it can prevent the stage or sensor from falling.

  • Camera Screw Adapter LV-0019

    Camera Screw Adapter LV-0019

    This is an adapter for mounting the LV-1800 sensor onto a large tripod head (1/4 inch screw).

  • large tripod

    large tripod

    It is used for installing sensors and stands in places without a base plate. Please use it together with the LV-0018A steel plate and the LV-0019 camera screw adapter.

  • Storage Trunk LV-0350

    Storage Trunk LV-0350

    This is a trunk that can store the main unit of the LV-1800 and options such as a large magnet stand.
    Items that can be stored
    ・LV-1800 × 1 / ・LV-0030 (+LV-0015/0016) × 1 / ・Objective lens × 2 / ・LV-0185 × 1 / ・LV-0018A × 1

  • Liquid measurement jig (made to order)

    Liquid measurement jig (made to order)

    This is a jig for efficiently irradiating and receiving laser light inside a liquid. Two types are prepared: a tube type and a quartz rod type compatible with solvents. You can choose to directly attach it to the sensor or have the jig suspended separately.
    ※ For details, please consult separately.