Micro-object vibration and measurement system

The micro-object vibration and measurement system enables micro-spot laser light and image observation using incident illumination by equipping the LV-1800 with the built-in camera LV-0181 for positioning, the LV-0185 lighting unit, and various objective lenses.
It is a system that allows laser positioning on micro-structures and detection of multiple points.

Features

  • The laser spot diameter can be focused to 2.5 µm or less (when using the LV-0152B model).

  • Using an internal digital camera and incident lighting, it is possible to measure the irradiation conditions of the laser beam.

  • Measurements can be taken from 0.3 Hz to 3 MHz according to standard specifications.

Specification

  Model name Product Name
1 LV-1800 Laser Doppler Vibrometer
2 LV-0181 Built-in camera for positioning
3 LV-0150B
LV-0151B
LV-0152B
Objective lenses (please refer to the table below for details)
4 LV-0185 lighting unit
5 LV-0030 large magnet stand
6 LV-0015 XY axis micro-movement stage
7 LV-0016 Z-axis micro-motion stage

A PC is required for image display.

objective lens

  LV-0150B LV-0151B LV-0152B
Objective Lens-no01 Objective Lens-no02 Objective Lens-no03
Specification Magnification: 5x
WD: 36.1 mm
Spot diameter: φ4 µm or less
Includes conversion adapter
Magnification: 10x
WD: 38.8 mm
Spot diameter: φ3 µm or less
Includes conversion adapter
Magnification: 20x
WD: 22.5 mm
Spot diameter: φ2.5 µm or less
Conversion adapter included
LV-0181 Observation Image Objective Lens-no04 Objective Lens-no05 Objective Lens-no06

*Target scale: 100 µm / graduation