High-speed F/V converter
FV-1500
The FV-1500 high-speed F/V converter is a highly responsive frequency (F)/voltage (V) converter that converts frequency signals proportional to rotational or moving speeds into voltage signals.
This system can handle the high-speed operation of motors used in hybrid and electric vehicles, as well as manufacturing lines in production plants. It can also measure startup characteristics, minute rotational fluctuations during steady-state rotation, and transient phenomena of flow velocity changes with high-speed response.
Features
-
Wide bandwidth 0.2 Hz to 320 kHz (1 Hz to 120 kHz for conventional products)
-
High-speed response: 1 cycle + 3.5 μs (1 cycle + 7.6 μs for conventional products)
-
Rapid deceleration tracking function that provides near-ideal output even when the frequency drops instantaneously.
-
Center frequency tracking function that amplifies only transient fluctuation components (optional)
-
Supports 2-phase signal input, enabling forward and reverse output based on rotation direction detection.
-
Comparator output (optional) for alarm control, etc.
Main uses
-
Motor startup characteristics
-
Measurement of rotational performance of various rotating equipment
-
Measurement of rotational fluctuations in engines and motors
-
Rotation performance test under overload
What is the rapid deceleration follow function?
This function slows down and stops the rotation output (analog speed) when the input signal decelerates rapidly, such as in a rotating object, and the input is no longer than the previous pulse interval.
When the tracking function is ON:
The analog output becomes 0 if no signal is input after waiting for the set number of preceding signal pulse intervals (1.5 to 16).
When tracking function is OFF:
After the input signal becomes 0 (OFF), the analog output is held at the minimum measurement frequency (0.2 Hz) before becoming 0.
Specification
| Input specifications | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Input terminal | BNC (CO2 type), terminal block | ||||||||
| Input format | Single-phase AC/DC voltage-free (open collector compatible +12V pull-up) switchable 90° phase difference 2-phase signal (DC input only) |
||||||||
| Input voltage | AC input signal voltage range: 0.3 Vp-p to 30 Vp-p | ||||||||
| DC input signal voltage range: Lo; +1 V or less, Hi; +4 to 30 V | |||||||||
| Filter |
OFF / 20 kHz / 120 kHz Low-Pass Filter |
||||||||
| Input Impedance | 2.2 kΩ | ||||||||
| AC coupling frequency | 0.9 Hz approx. -3 dB | ||||||||
| Minimum input pulse width | 1.5 μs | ||||||||
| Input signal edge switching | Standing Up/Standing Down Switching | ||||||||
| Input frequency range | 0.2 Hz to 320 kHz
|
||||||||
| Signal output specifications | |||||||||
| Analog output signal | |||||||||
| Analog output terminals |
BNC/C02 type (voltage output) or |
||||||||
| Conversion time | 1 cycle within +3.5μs | ||||||||
| D/A resolution |
16 bit |
||||||||
| Filter |
OFF / 3 Hz / 10 Hz / 1 kHz switchable low-pass filter |
||||||||
| Analog output terminal signal |
|
||||||||
| Output DC offset | Voltage output: within ±10 mV | ||||||||
| Current output: ± 0.01 mA or less | |||||||||
| Temperature coefficient | Voltage output: ± 0.02 %/FS/℃ | ||||||||
| Current output: ± 0.07 %/FS/℃ | |||||||||
| Output Impedance | 50 Ω or more (voltage output) | ||||||||
| Linear number (In full-scale mode) |
|
||||||||
| pulse output signal | |||||||||
| Pulse output | BNC (CO2 type) Output terminal voltage: Lo: +1 V or less, Hi: +4.5 V or more Load resistance: 100 kΩ or more |
||||||||
| Pulse output option (FV-0154) |
Open collector output Withstand voltage: DC 30 V, sink current: 50 mA or less |
||||||||
| Comparator output signal (optional FV-0152) | |||||||||
| Comparator output terminal | Terminal block (Phoenix Contact MC1,5/6-STF-3,81) | ||||||||
| Output format | Semiconductor relay (1 make contact) 2CH | ||||||||
| Maximum contact capacity | DC 30 V / 1 A | ||||||||
| Contact ON resistance | 25 Ω or less | ||||||||
| setting | UPPER, LOWER Relay turns ON if UPPER setting value ≤ measured value, and ON if LOWER setting value > measured value. |
||||||||
| Additional functions | Delay function | The contact can be turned ON or OFF (0 to 100 times) if the set threshold is exceeded for a set number of consecutive times. | |||||||
| Hysteresis function | Change the return frequency or rotation speed, or movement speed by 0-20%. Adjustable in 1% increments |
||||||||
| Hold function | Change the minimum output time for holding the contact ON state from 1 ms to 1000 ms. Adjustable in 1 ms increments |
||||||||
| Function | |||||||||
| moving average | OFF / Simple average 2 to 720 times / Exponential average 2 to 64 times | ||||||||
| Number of pulses | 1 to 999999 P/R | ||||||||
| factor | 0.0001~99999 | ||||||||
| Multiplication/Divide | Multiplication: OFF / 2 / 4 (only when inputting 2-phase signals with 90° phase difference) Frequency division: OFF, 2 to 1000 | ||||||||
| Condition Memory | Remember 10 conditions | ||||||||
| Rapid deceleration tracking (predictive calculation) | OFF / TYPE1 / TYPE2 | ||||||||
| Estimated stopping time | OFF / NUM: 1.5x - 16x / TIME: 1 ms - 4000 ms | ||||||||
| Direction determination | OFF / ON (only when inputting 2-phase signal with 90° phase difference) | ||||||||
| CAL | Voltage output: ZERO: 0 V, FULL: 10 V | ||||||||
| Current output: ZERO: 0 mA (4 mA), FULL: 16 mA (20 mA) | |||||||||
| Display | |||||||||
| Display accuracy | Displayed value* X (±0.01%) ± 1 count (*The displayed value is the count value excluding the decimal point) | ||||||||
| Display | Fluorescent display tube (display area 69.85 mm × 11.45 mm), 4-level brightness adjustment. | ||||||||
| Display update time | 0.5s, 1s | ||||||||
| Display unit * | Hz, r/min, m/min, USER | ||||||||
| Status LED display | SIG | It lights up each time a signal is input. | |||||||
| Measurement mode | Full scale, deviation, automatic deviation tracking (optional FV-0151) | ||||||||
| Power supply for sensors | |||||||||
| Power supply for sensors | Selectable between +12V ± 10% 150mA and +5V ± 10% 150mA via the rear switch. | ||||||||
| Applicable standards | |||||||||
| CE marking | Low Voltage Directive EN61010-1:2010 | ||||||||
| EMC Directive EN61326-1:2013 | |||||||||
| General specifications | |||||||||
| Input/Output Ground | The signal grounds for the signal outputs (V OUT, I OUT) are common (connected to the chassis). The signal grounds for the signal inputs (SIG IN, SIG 1, SIG 2) and the sensor power supply (+12V, +5V) are common. The ground for signal inputs and sensor power supplies is isolated from the enclosure. The ground between inputs and outputs is isolated. The withstand voltage between isolated signal grounds and the chassis is up to 42Vpk |
||||||||
| Power supply voltage range | Includes a dedicated DC 16V AC adapter (AC 100-240V). | ||||||||
| Power consumption | Approximately 36 VA or less | ||||||||
| Operating temperature range | 0 to 40 degrees Celsius | ||||||||
| Storage temperature range | -10 to 50°C | ||||||||
| Operating humidity range | 5-80% (but no condensation) | ||||||||
| Storage humidity range | 5-85% (but no condensation) | ||||||||
| Outline drawings | 210(W)×44(D)×200(D) mm (excluding protrusions)
|
||||||||
| Weight | Approximately 1 kg | ||||||||
| Accessories | Dedicated AC adapter (PS-P20023 / VM1391-VM1700 2M) x 1, Instruction manual x 1, Connector (MC1.5/6-STF-3.81) x 1 |
||||||||
| Price/Options | |||||||||
| price | |||||||||
| option |
FV-0151 Automatic center frequency tracking function | ||||||||
| FV-0152 Comparator Output Function | |||||||||
| FV-0153 Output voltage deviation change (±10 V) | |||||||||
| FV-0154 Open collector output function (pulse output) | |||||||||
| FV-0014 Panel mounting bracket | |||||||||
*When measuring travel speed (m/min) in combination with the RP-7400 1200 P/R, please set the display unit to "USER".