Skip to content

Select your region & language

Global

Region

Sound Level Meter
LA-1441A/LA-1411/LA-4441A

LA-1411
LA-1441A
LA-4441A
LA-4441A
LA-4441A
LA-4441A

Integrating average sound level meter (noise meter) "LA-1411/LA-1441A (Class 2, standard sound level meter)" and
The three "LA-4441A (Class 1, precision sound level meter)" products comply with the Measurement Law (enacted in 2021), IEC 61672-1:2013, and JIS C 1509-1:2017. *1
With a wide linearity range of 100 dB, it can simultaneously measure, select, and display seven noise indicators, including time-averaged sound level, in response to a wide range of sound volume fluctuations without switching level ranges.
These three products allow you to separate the microphone unit from the main unit of the sound level meter and connect a dedicated coaxial cable to extend the microphone unit over long distances.

With a curved body that minimizes the effects of sound reflection, and an easy-to-read and intuitive operation menu, we have achieved a more user-friendly human interface than ever before, in addition to improved functionality.

Furthermore, the LA-1441A and LA-4441A sound level meters (noise meters) allow for quick and easy data transfer to a PC via RS-232C or USB using dedicated free software, and also enable remote operation and control of the sound meter from a PC.

ASNITE and JCSS calibration are available.

*1 Starting with shipments in December 2017, the compliance standards listed on the back of the unit have been changed to IEC 61672-1:2013 and JIS C 1509-1:2017. Shipments prior to November 2017 have also been confirmed to comply with IEC 61672-1:2013 and JIS C 1509-1:2017.
Customers who purchased our products before November 2017 and wish to have their JIS/IEC markings updated during calibration (for a fee) should contact our customer service center.

LA-1411
Class 2 (Standard Sound Level Meter)
  • Sound level meters for machine noise levels and product inspection.
  • Low-priced, high-performance models
LA-1441A
Class 2 (Standard Sound Level Meter)
  • Sound level meter for measuring noise in work environments and ambient noise.
  • Equipped with automatic measurement functions using timers and triggers.
  • Level trend recording is possible with a saving interval of 1 ms (L p record function).
  • Equipped with USB and RS-232C interfaces.
LA-4441A
Class 1 (Precision Sound Level Meter)
  • Sound level meter for noise analysis and product inspection.
  • Equipped with a 10 ms IMPULSE time weighting characteristic to handle rapidly fluctuating sound levels.
  • Level trend recording is possible with a saving interval of 1 ms (L p record function).
  • Equipped with USB and RS-232C interfaces.

Features

Simultaneously measures and displays seven noise indices, including time-averaged sound level (Leq).

Equipped with a time-averaged sound level (Leq) measurement function, which is essential for noise measurement.
It simultaneously measures seven noise indices (L p, L eq, L E, L max, L min, L peak, L N <L high, L 5, L 10, L 50, L 90, L 95, L low, Lave >), allowing for selection and display, and preventing measurement omissions.

L p Time-weighted sound levels
L eq Time-averaged sound level
L E Acoustic exposure level
L Max Maximum value of time-weighted sound level
L min Minimum time-weighted sound level
L Peak Sound Level
L N Time-based sound level

The microphone and main unit of the sound level meter can be separated and extended over long distances using a dedicated coaxial cable.

The microphone and main unit of the sound level meter can be separated and extended over long distances using a dedicated coaxial cable. This makes remote measurements easy.

Easy and convenient data processing

The sound level meter can be connected to a Windows® PC via RS-232C or USB cable, and using dedicated free software, acquired measurement data can be quickly and easily transferred to the PC, and the sound level meter can be remotely operated and controlled from the PC.
(Excluding LA-1411)

  • Easy and convenient data processing

System Configurations

img-a1411-1441a-4441a-02

AX-501 AG-2000 AG-2000

Specification

Product Specifications

  LA-1411 LA-1441A LA-4441A
Applicable standards JIS C 1509-1:2017 (Class 2)
JIS C 1516:2020 (Class 2)
JIS C 1509-1:2017 (Class 1)
JIS C 1516:2020 (Class 1)
IEC 61672-1:2013 Class 2 IEC 61672-1:2013 Class 1
Type Approval Number
Measurement Law (enforced in 2021)
TS223 TF226
Microphone 1/2 Inchback electret
Condenser microphone
MI-1433

Nominal sensitivity level: -29 dB
1/2 Inchback electret
Condenser microphone
MI-1235

Nominal sensitivity level: -29 dB
Preamplifier Microphone preamplifier MI-3111
Linear motion range
(Linearity range)
100 dB (wide range), 80 dB (normal range)
Measurement frequency range 10 Hz to 8 kHz (JIS, IEC) 10 Hz to 20 kHz (JIS, IEC)
Full range of linear motion A-weighting: 26 dB to 135 dB (JIS, IEC)
C characteristics: 31 dB to 135 dB (JIS, IEC)
Z characteristic: 36 dB to 135 dB (JIS, IEC)
A-weighting: 27 dB to 135 dB (JIS, IEC)
C characteristic: 30 dB to 135 dB (JIS, IEC)
Z characteristic: 35 dB to 135 dB (JIS, IEC)
Frequency weighting characteristics A, C, Z
Self-noise A-weighting: 20 dB or less
C characteristic: 25 dB or less
Z characteristic: 30 dB or less
A-weighting: 19 dB or less
C characteristic: 22 dB or less
Z characteristic: 27 dB or less
Level range 20-90 dB, 30-90 dB, 40-100 dB, 50-110 dB, 60-120 dB, 70-130 dB (normal range) /
40–120 dB (wide range)
Time-weighted properties FAST,
SLOW
FAST,
SLOW,
Impulse,
10 ms
Measurement items L p, L eq, L E, L max, L min, L peak, L N (L high, L 5, L 10, L 50, L 90, L 95, L low, L ave)
Level Trend L p, various calculation values L p, various calculated values
(Displayed values can be read via auto-memory recall)
Sampling period 31.25 μs (other than L N)
100ms (L N)
20.83 μs (other than L N)
100ms (L N)
Measurement time Manual, arbitrary setting (199 h 59 m 59 s max)
Total measurement time --- Arbitrary setting (199 h 59 m 59 s max)
Start mode Manual Manual, Timer (00 h:00 m (1 min units)
trigger
(After trigger detection, it can be started after any time has elapsed (0 to 10 seconds, in 1-second increments) (trigger delay function))
Internal reference signal (REF OUT) 1 kHz sine wave (relative to the upper limit of the level range: -6 dB (normal range), -16 dB (wide range))
Recommended acoustic calibrator SC-2600, SC-2500A, SC-2120A SC-2600, SC-2500A
Display LED-backlit transflective LCD, 124 x 64 dot matrix display.
Numerical and bar indicator display of measured values,
List display of various calculated values and trend graph display,
Date and time, measurement time, and status display of various devices.
Display update time Numerical display: 1 s, Bar indicator: 0.1 s, Trend screen: 0.2 s
Manual memory function Total calculation values: 100 data sets
Auto memory function --- L eq: 1000 data × 200 blocks,
Total calculation values: 100 data sets × 200 blocks
LP record
Memory function
--- Time-weighted sound level (LP) at specified time intervals
Save (1000 data × 200 blocks)
Select from the following save intervals: 1 ms, 2 ms, 5 ms, 10 ms.
Measurement condition memory function Condition memory (9 entries)
Resume function The power-on settings can be set to the last shutdown settings, factory default settings, or
Can be set to any of the specified condition memory numbers.
Backup function

The built-in rechargeable battery allows for backup of the clock and data stored in memory.

  • Charging time to full charge: Approximately 24 hours
  • Backup period: 3 months or more (provided the data is stored at 23°C).
AC output *1 At the upper limit of the bar indicator:
0.707 Vrms/FS (Normal Range)
2.236 Vrms/FS (Wide Range)
Load resistance 10 kΩ or more
*Frequency weighting can be selected from either linked to settings or fixed to Z characteristic.
DC output *1 At the upper limit of the bar indicator: 2.5 V/FS,
Voltage change per 10 dB: 0.25 V,
Load resistance 10 kΩ or more
External control input --- Action: Toggle start/pause function
Hardware specifications: Negative logic pulse input
(Voltage levels: Low: 0.0–0.5 V, Hi: 3.0–5.0 V)
Pulse width: 100 ms or more, Absolute maximum rated input voltage: 5 V
Comparator function
(Option: LA-0141) *2
--- Setting comparison range: 40.0 to 130.0 dB (in 0.1 dB increments)
Comparison item: L max or L peak
Delay time: 0 ms, 100 ms, 1 s, 5 s
HOLD time: 100ms, 1s, 5s, manual
Output logic: Negative logic or positive logic
Output format: Open collector method
(Applicable voltage: DC +3 to +24 V, current 50 mA or less)
Clock function Built-in (Year/Month/Day Hour:Minute)
RS-232C --- Control of the main unit and data output, output to the printer (DPU-414)
Transfer speed: 9,600/115,200 bps
USB --- Control and data output of the main unit,
USB 2.0 FULL SPEED (Virtual COM port connection) * 6
Cable extension *3
(Between microphone and main unit)
MAX. 305 m MAX. 105 m
Printer function --- Contents: Measured values, start/end times of measurement, actual measurement time, frequency weighting characteristics, time weighting characteristics
Mode: Normal (alphanumeric) or screen image; Compatible printer: DPU-414
Power supply Two AA batteries
Alternatively, PB-7090 AC adapter (optional)
Battery life
(continuous use)
Approximately 8 hours
(Using LR6 alkaline batteries, 23°C/backlight off/RS-232C and USB not in use)
Operating (storage) temperature range -10 to 50ºC (-20 to 60ºC)
Humidity range for use (storage) 25 to 90% RH (10 to 90% RH) (non-condensing)
Size Approximately 78 (W) × 263 (H) × 47 (D) mm
mass
(Including batteries)
Approximately 360 g (including batteries)
Accessories Windscreen (⌀ 70 mm), carrying case, hand strap, batteries, instruction CD *7, quick start guide
price
(tax included)
     
Inspection fee at time of shipment
(Tax included) *4
   
Option
  • LA-0141 コンパレーター出力オプション:
  • SC-2600 音響校正器(クラス 1):
  • SC-2500A 音響校正器(クラス 1):
  • SC-2120A 音響校正器(クラス 2):
  • AG-2010 マイクロホン延長ケーブル(10 m):
  • AG-2030 マイクロホン延長ケーブル(30 m):
  • AG-2050 同上コードリール付き(50 m):
  • AG-2100A 同上コードリール付き(100 m):
  • PB-7090 ACアダプター(AC 100~240 V) ※5
  • AX-501 信号入出力ケーブル(2 m):
  • AX-5022B RS-232C ケーブル(2 m):

*1 AC output/DC output can be switched using the sound level meter's menu.
*2 The comparator function (optional: LA-0141) cannot be installed on the LA-1411.
*3 The cable extension distances listed here are based on using our company's dedicated cables. CE marking compliance applies to distances of less than 30 m.
*4 As a general rule, our company replaces the inspection conducted at the time of shipment with an "Approval Mark" obtained through our own inspection.
After shipment, we inspect and adjust our products, and then request inspection from the Japan Quality Assurance Organization (JQA).
Furthermore, since our standard product is one without certification, customers who require a product with certification (certificate of conformity mark) or who wish to have their product certified by a public institution from the time of shipment should inform us when placing their order.
*5 The included AC cord is for use in Japan. If you wish to use it with overseas specifications (AC 120-240V), please specify the appropriate AC cord separately.
*6 For sample programs and utilities, please refer to the Free Utility Program Downloads.
*7 The LA-1411 comes with an instruction manual and a quick start guide (an instruction CD is not included).

Accessories

  • List of accessories
    List of accessories
  • Included carrying case
    Included carrying case

Option

LA-0141 Comparator Output Option
Overview This option outputs a signal when a set value is exceeded. It can be used to determine the sound level of the object being measured.
Specification
Setting comparison value : 40.0–130.0 dB (in 0.1 dB increments)
Items to compare : L max or L peak
Delay time : 0 ms, 100 ms, 1 s, 5 s
(This is useful when you only want to make a judgment when the set value is exceeded consecutively. It helps avoid false judgments caused by single noise events, etc. Note that this is ineffective when the HOLD setting is manual.)
HOLD time :100ms, 1s, 5s, manual
(This is the time the output is maintained even if the input falls below the set value. In manual mode, it is maintained until the next measurement starts.)
Output logic : Negative logic or positive logic
Output format Open collector type (applied voltage: DC +3 to +24 V, current 50 mA or less)
Price (tax included)  
Recommended Sound Calibrators (SC-2600, SC-2500A / SC-2120A)
Overview This acoustic calibrator is essential for noise measurement. It conforms to JIS C1515/IEC 60942 standards.
View details
SC-2600
SC-2600
SC-2500A
SC-2500A
SC-2120A
SC-2120A
price
  • SC-2600 Acoustic Calibrator (Class 1):
     
  • SC-2500A Acoustic Calibrator (Class 1):
     
  • SC-2120A Simple Acoustic Calibrator (Class 2):
     

LA-0207A + LA-0208A All-Weather Windproof Screen

Overview
Using an all-weather windproof screen can reduce wind noise during outdoor noise measurements and prevent microphone damage from rain, snow, etc.

Screen diameter: 200 mm
Mounting screw thread: 1/4-20 UNS depth 6 (camera tripod screw)

*With LA-0207A + LA-0208A installed,
Conforms to JIS C 1509-1:2017 / IEC 61672-1:2013
*Consumable parts (windscreen, waterproof mesh) will need to be replaced periodically.

LA-0207A + LA-0208A All-Weather Windproof Screen

price
  • LA-0207A All-weather windproof screen:
     
  • LA-0208A AG-2000 dedicated adapter:
     

AG-2000 Series Microphone Extension Cable

Overview The microphone unit and the main unit can be separated, and the microphone can be extended over long distances using a dedicated microphone extension cable.
Multi-core cables tend to be heavy, but dedicated microphone extension cables use coaxial cables, so even with a 100m extension, the cord reel can be small, making it easy to move around on site.

AG-2000 Series Microphone Extension Cable

price
  • AG-2010 Microphone Extension Cable (10 m):
     
  • AG-2030 Microphone Extension Cable (30 m):
     
  • AG-2050 Same as above, with cord reel (50 m):
     
  • AG-2100A Same as above, with cord reel (100 m):
     
  • MI-0301 comes with a microphone holder.
  • The extension cable with a cord reel includes a 5-meter connecting cable for the sound level meter unit.
  • You can extend the cable length by connecting multiple extension cables.

AX-501 Output Cable

Overview This is for AC/DC signal output, comparator output, and external control signal input.
This cable is used for connecting to FFT analyzers, data recorders, PCs, PLCs, etc. (2 m)

AX-501 Output Cable

price  

AX-5022B RS-232C cable

Overview This is an IF cable that connects the product to a PC (RS-232C connector).
This cable is used for connecting to a computer. (2 m)

AX-5022B RS-232C cable

price  

Tripod for sound level meter

Overview This is a standard tripod for sound level meters. (Slik Airy L100)
Tripod for sound level meter

PB-7090 AC Adapter

Overview
Input voltage: AC 100-240 V, 50/60 Hz
Output voltage: DC 5.9 V
Output connector: EIAJ RC-5320A Voltage Class 2
Rated output current: 3.5 A
Total length: 3.5 m
*The included AC cord is for use in Japan. If you wish to use it with overseas specifications (AC 120-240V), please specify a compatible AC cord separately.
PB-7090 AC Adapter
price  

Calibration

To guarantee the accuracy of the values measured by a measuring instrument, the instrument must be calibrated at appropriate intervals. Only by using a calibrated instrument can the reliability and stability of the obtained measurements be improved. There are three calibration services available for the integral-averaged sound level meter: "ASNITE calibration," "JCSS calibration," and "general calibration."