Skip to content

Select your region & language

Global

Region

Measurement Microphone and Preamplifier
MIシリーズ

A wide range of products to meet your needs for measuring sound.

The MI series
A sound field type measurement microphone employing a back electret design.
● Microphone for measurement
● Microphone preamplifier
● TEDS microphone for measurement
● Low-noise microphone
We have the following available.

Features

MI-1271/MI-3170
Microphones and preamplifiers for measurement

Microphone

Features

MI-1271
MI-1271

The MI-1271 is a precision-grade 1/2-inch back-electret microphone.
By employing a laser welding fixing method for the titanium vibrating membrane, we have achieved extremely high stability.
In addition, its wide temperature and humidity range (-30 °C to 80 °C / 0 to 90 %RH) and wide frequency range (1 Hz to 20 kHz) make it a high-performance microphone capable of meeting a wide range of applications, such as measuring ambient noise in hot and humid environments, measuring ultra-low frequency sounds (up to 20 Hz), and measuring sound during cold-start tests of automobiles.

MI-1235
MI-1235

The MI-1235 has a frequency range that covers the audible range from 10 Hz to 20 kHz, making it ideal for measuring and analyzing ambient noise and building acoustics, as well as for measuring and analyzing radiated sound from electronic and mechanical equipment.
This is a precision-class microphone designed to be used in combination with the MI-3111 microphone preamplifier.

MI-1433
MI-1433

The MI-1433 has a frequency range of 20 Hz to 8 kHz, making it ideal for applications such as measuring ambient noise. It is a standard, low-cost, general-purpose microphone designed to be used in combination with the MI-3111 microphone preamplifier.

MI-1531
MI-1531

The MI-1531 is a 1/4-inch back-electret microphone with a wide frequency range of 10 Hz to 100 kHz. It is ideal for measurements in confined spaces. When combined with the MI-3140 preamplifier, it can measure down to 157 dB.

Free-field response

  • Free-field response

Microphone specifications

Model name
MI-1271
MI-1235
MI-1433
MI-1531
size
1/2 inch
1/4 inch
Polarization voltage
0 V
Sensitivity
-26 dB±1.5 dB re. 1 V/Pa
50 mV/Pa (1 kHz)
-29 dB ±3 dB re. 1 V/Pa
36 mV/Pa (1 kHz)
-48 dB ±3 dB re. 1 V/Pa
4 mV/Pa (250 Hz)
Frequency range
1 Hz to 20 kHz
10 Hz to 20 kHz
20 Hz to 8 kHz
10 Hz to 100 kHz
(Without protective grid)
10 Hz to 20 kHz
(With protective grid)
Maximum sound pressure level
135 dB (when using MI-3170)
135 dB (when using MI-3111)
157 dB (when using MI-3140)
Self-noise level
(A characteristic)
14.0 dB (typical value)
(When using MI-3170)
19.0 dB (typical value)
(When using MI-3111)
37 dB (typical value)
(When using MI-3140)
Operating temperature range
-30 to +80 °C
-10 to +50 °C
-30 to +60 °C (when using MI-3140)
Operating humidity range
0-90% RH (non-condensing)
20-90% RH (non-condensing)
0-90% RH (non-condensing)
Storage temperature range
-40 to +70°C
-20 to +60 °C
-30 to +80 °C
Storage humidity range
0-90% RH (non-condensing)
10-90% RH (non-condensing)
0-90% RH (non-condensing)
Outline drawings
⌀13.2 mm × 16.9 mm
⌀13.2 mm × 13.7 mm
⌀13.2 mm × 13.5 mm
⌀7.0 mm × 10.5 mm
Weight
Approximately 6 g
Approximately 1.5 g
Compatible preamplifiers
MI-3170
MI-3111
MI-3140
Price (tax included)
 
 
 
 

Preamplifier

Features

MI-3170
MI-3170

The MI-3170 is a preamplifier for 1/2-inch back-electret condenser microphones and can be used in combination with the MI-1271 microphone. Its CCLD (constant current drive) design and BNC connector allow for direct connection to numerous CCLD-compatible amplifiers, analysis equipment, data recorders, and other devices. When combined with the MI-1271 high-performance microphone, it covers an ultra-low frequency measurement range from 1 Hz and enables measurements over a wide operating temperature range (-30 °C to 80 °C).

MI-3111
MI-3111

The MI-3111 is a preamplifier for 1/2-inch back-electret condenser microphones. It uses a CCLD (constant current drive) system and has a BNC connector.
This is a general-purpose preamplifier that can be used in combination with microphones such as the MI-1235/MI-1433.

MI-3140
MI-3140

The MI-3140 is a 1/4-inch back-electret condenser microphone preamplifier. It is ideal for space-constrained measurements. It features CCLD (constant current drive) operation and a 10-32 UNF connector.
This is a dedicated preamplifier for use with the MI-1531 microphone.

Preamplifier specifications

Model name
MI-3170
MI-3111
MI-3140
size
1/2 inch
1/4 inch
Compatible microphones
MI-1271/1235/1433
MI-1235/1433
MI-1531
Insertion loss
0.15 dB (at 1 kHz) (typical value)
1.0 dB (typical value)
0.35 dB (typical value)
Frequency range
10 Hz to 40 kHz (+0.1 dB, -0.2 dB, 1 kHz reference)
1 Hz to 40 kHz (+0.1 dB, -1.5 dB, 10 Hz reference)
10 Hz to 20 kHz (±1.0 dB, 1 kHz reference)
20 Hz to 20 kHz (±0.6 dB, 1 kHz reference)
10 Hz to 100 kHz (within ±0.5 dB, 1 kHz reference)
(When used in combination with the SR-2210 2ch sensor amplifier, the frequency range is up to 20 kHz.)
Self-noise (RMS voltage) A-weighted
3.3 μV or less
5.0 μV or less
2.5 μV or less (20 Hz to 20 kHz)
Maximum output voltage
±8 V (peak value) (Note 1)
Sound pressure equivalent to 135 dB
(When using a microphone with a sensitivity of 50 mV/Pa)
±5.6 V (peak value) (Note 1)
Sound pressure equivalent to 135 dB
(When using a microphone with a sensitivity of 35 mV/Pa)
±8 V (at DC 24 V)
Operating temperature range
-30 to +80 °C
-10 to +50 °C
-30 to +60 °C *
Operating humidity range
0-90% RH (non-condensing)
30-90% RH (non-condensing)
0-90% RH (non-condensing)
Storage temperature range
-40 to +70°C
-20 to +60 °C
-30 to +80 °C
Storage humidity range
0-90% RH (non-condensing)
10-90% RH (non-condensing)
0 to 95% RH (non-condensing)
Power supply
CCLD (Constant Current Drive)
2–4.5 mA, DC 18–26 V
CCLD (Constant Current Drive)
0.5 to 5 mA, DC 15 to 25 V
CCLD (Constant Current Drive)
2 to 20 mA, DC 18 to 25 V
Compatible connectors
C02(BNC)
10-32 UNF
Outline drawings
φ12.7 mm × 80.5 mm
φ12.7 mm × 63.5 mm
φ6.35 mm × 44 mm
Weight
Approximately 35 g (excluding microphone)
Approximately 25 g (excluding microphone)
Approximately 5.5 g (excluding microphone)
Signal cable
MX-1000 series (recommended), MX-2000 series (recommended)
MX-2000 series (recommended)
NP-0120 series (recommended), NP-0180 series (recommended)
Accessories
Input connector protective cap × 1
Preamplifier holder (MI-0301) x 1
Instruction manual × 1
Input connector protective cap × 1
Preamplifier holder (MI-0301) x 1
Instruction manual × 1
Input connector protective cap × 1
1/4-inch conversion adapter (SC-0313) × 1
Instruction manual × 1
Price (tax included)
 
 
 

*This is the operating temperature range for the main unit only. The operating temperature range including the NP-0120 series signal cable is -25 to 60 °C.

TEDS Microphone MI-1271M12 for Measurement

TEDS microphone MI-1271M12 for measurement
MI-1271M12

The MI-1271M12 TEDS microphone for measurement is a microphone that integrates the MI-1271 microphone and the MI-3170 preamplifier into a single unit.

This device can be used by directly connecting it to equipment equipped with a CCLD (Constant Current Drive) power supply via a BNC cable. When connected to TEDS-compatible equipment, information such as sensitivity is automatically read. As a result, it enables reliable measurement and time savings.

Features

  • TEDS compatible

  • By employing a titanium diaphragm and improving electret performance, long-term stability and a wide frequency range from 1 Hz to 20 kHz are achieved.

  • It can be used in a wide temperature range from -30°C to +80°C.

  • It can be used by directly connecting it to equipment equipped with a CCLD (Constant Current Drive) power supply via a BNC cable.

MI-1271M12 Specifications

MI-1271M12
Specification
Size, response type, polarization voltage
1/2 inch, free-field type, 0V
Sensitivity
-26.0 dB ± 1.5 dB re.1 V/Pa (50 mV/Pa)
Frequency range
1 Hz to 20 kHz (±2 dB)
Maximum sound pressure level (total harmonic distortion 3%, 250 Hz)
135 dB
Self-noise level (A-weighted)
14.0 dB (typical value)
Static pressure characteristics (250 Hz)
-0.013 dB/kPa
Temperature characteristics (250 Hz)
+0.005 dB/K
Humidity characteristics (250 Hz)
-0.0004 dB/%
Operating temperature range
-30 to +80 °C
Operating humidity range
0 to 90% RH (non-condensing)
Storage temperature range
-40 to +70°C
Storage humidity range
0 to 90% RH (non-condensing)
Power supply
CCLD (Constant Current Drive)
Drive current
2 to 4.5 mA (rated 4 mA)
Power supply voltage
DC18V to 26V (rated 24V)
Output connector
C02(BNC)
TEDS Version**
IEEE 1451.4: 2004 (Template: Microphone with built-in Preamplifier Ver.1.0)
Outline drawings
⌀13.2 x 91.9 mm
Weight
Approx. 41g
Accessories
Proofreading chart: 1 copy
Instruction manual: 1 copy
Preamplifier holder (MI-0301): 1 piece
Price (tax included)
 
  • Reference environmental conditions: 23 °C, 50%RH, 101.3 kPa

  • The cable length must be 30m or less. If extended beyond this length, it will not comply with CE marking. Please be aware of this.

  • For information regarding TEDS compatibility of the measuring instruments and amplifiers you wish to connect, please contact your retailer or our company.

Low-noise microphone MI-1282M10

MI-1282M10
MI-1282M10

To create a more comfortable sound environment, there is a growing trend towards reducing noise levels in products such as home appliances and information equipment, as well as the sounds of motors and inverters that penetrate the cabin of vehicles. Low-noise microphones enable measurements of 10 dB (A-weighted) or less, which is difficult with conventional microphones.
Conventional low-noise microphones required a separate amplifier for power supply, but this product employs a CCLD (constant current drive) method, allowing it to be directly connected to a CCLD-compatible analyzer.

Features

  • Self-noise level: 4.5 dB (typical value)

  • Measurement is possible by directly connecting a BNC cable to CCLD (constant current drive) compatible equipment.
    Using the DS-5000 with battery power eliminates the need for a separate power source.

  • TEDS compatible

MI-1282M10 Specifications

MI-1282M10
Specification
Bore diameter, response type, polarization voltage
1/2 inch, free-field type, 0V
Sensitivity
-6.0 dB ± 2.0 dB re.1V/Pa (500 mV/Pa)
Frequency range
10 Hz to 20 kHz (±3 dB)
10 Hz to 16 kHz (±2 dB)
Maximum sound pressure level (1 kHz, total harmonic distortion 3%)
>100 dB (10 Hz to 5 kHz)
>80 dB (5 kHz to 20 kHz)
Self-noise level (A-weighted)
4.5 dB (typical value)
Operating temperature range
-10 to +60 °C
Operating humidity range
0 to 85% RH (non-condensing)
Storage temperature range
-20 to +70 °C
Storage humidity range
0 to 85% RH (non-condensing)
Power supply
CCLD (Constant Current Drive)
Drive current
2.8 to 20 mA (rated 4 mA)
Power supply voltage
24 to 32 VDC (24 V rating)
Output connector
C02 (JIS C5412-1995 compliant BNC jack)
TEDS Version**
IEEE 1451.4: 2004 (Template: Microphone with built-in Preamplifier Ver.1.0)
Outline drawings
⌀13.2 mm x 91.2 mm
Weight
Approx. 37g
Accessories
Proofreading chart: 1 copy
Instruction manual: 1 copy
Preamplifier holder (MI-0301): 1 piece
Packaging case: 1
Price (tax included)
 
  • Reference environmental conditions: 23 °C, 50%RH, 101.3 kPa

  • The cable length must be 20m or less. If extended beyond this length, it will not comply with CE marking. Please be aware of this.

  • For information regarding TEDS compatibility of the measuring instruments and amplifiers you wish to connect, please contact your retailer or our company.

  • We do not guarantee operation when connected to the SR-2210.

Connection Configuration

Each product illustration in the diagram below links to its respective dedicated product page.

Connection Configuration

* For measurements up to 100 kHz, please consider the CF-9200A/9400A or DS-5000 100 kHz unit. (Discontinued products) The CF-7200, DS-3000, and DS-2000 series 100 kHz units are also compatible.

* MI-1282M10 Low Noise Microphone Compatible Models

  • This software is compatible with the FFT analyzers DS-5000 series, CF-9200A/CF-9400A, and CF-4700A.

  • When used in combination with the FFT analyzer DS-3000 series, CF-9200/CF-9400 (without A), or CF-4700 (without A), sufficient performance may not be achieved depending on the conditions.
    Modifications will be necessary when using these devices in combination. After modifications, some functions may be limited.
    For further details, please contact your nearest sales office.