| Maximum input charge |
±100000 pc |
| Input connector |
Miniature connector |
| Charge capacitor |
100 pF, 1000 pF, 10000 pF |
| Leakage resistance |
10 GΩ, 1 GΩ, 100 MΩ |
| Frequency response characteristics |
| acceleration |
1.0 Hz to 15 kHz ± 0.5 dB 0.2 Hz to 50 kHz ± 3 dB |
| speed |
3.0 Hz to 3 kHz ± 0.5 dB |
| displacement |
3.0 Hz to 500 Hz ± 1 dB |
(However, 160 Hz is considered 0 dB.) |
| accuracy |
| acceleration |
± 2% |
| speed |
± 3% |
| displacement |
± 5% |
(At 160Hz, 25℃±3℃) |
| Rated output voltage |
±10 V |
| Maximum output load |
3 mA, 1500 pF |
| Output connector |
C02 (BNC) |
| Input-referred noise level |
0.05 pC (rms) or less |
| Output offset |
±5mV or less |
| Sensitivity |
0.01–999 pC/EU |
| Filter |
HPF: through, 3 Hz, 10 Hz (-18 dB/oct) LPF: through, 1 kHz, 10 kHz (-18 dB/oct) (10Hz and 1kHz conform to "Vibration of rotating machinery - Requirements for vibration severity measurement" <JIS B 0907-1989>) |
| CAL signal (calibration) |
160 Hz ± 5%, 80 Hz ± 5% 1 V 0-p ± 2.0% Sine wave (25°C ± 3°C) |
| Maximum input warning display function |
OFF: If the output exceeds ±10 V, the [-10 dB/Over] indicator will flash red. When it returns to within ±10 V, the flashing will stop. ON: If the output exceeds ±10 V, the red light will continue to blink until the [▲/RST] switch is pressed. |
| Output sensitivity |
| acceleration |
0.01, 0.0316, 0.10, 0.316, 1.00, 3.16, 10.0, 31.6, 100, 316, 1000 mV/EU |
| speed |
0.01, 0.0316, 0.10, 0.316, 1.00, 3.16, 10.0, 31.6, 100, 316, 1000 mV/EU |
| displacement |
0.10, 0.316, 1.00, 3.16, 10.0, 31.6, 100, 316, 1000 mV/EU |
|
| Auto power save |
If there is no input from any switch for more than approximately 2 minutes, only the decimal point will be displayed on the main display. |
| Other functions |
- Output level indicator: Green light at -10 dB full scale, red blinking when over full scale.
- Condition Memory: Retains settings even when the power is turned off.
- Condition Lock: Prevents accidental key operation.
|
| Power supply voltage |
10V to 15VDC (AC 100V can be used with a separately sold AC adapter) |
| Current consumption |
120 mA or less (when 12 VDC IN displays "1.00") |
| Link |
Up to 6 units per AC adapter |
| Outline drawings |
28 (W) × 121 (H) × 194 (D) mm (excluding protrusions) |
| Weight |
Approximately 510g |
| Operating environment conditions |
-10°C to +50°C, 90% RH or less (non-condensing) |
| Storage environmental conditions |
-10°C to +50°C, 90% RH or less (non-condensing) |
| Accessories |
Joint cables, stabilizers, connecting fittings |
| price |
|
| Option |
- AC adapter PS-P20022C (PE1821078)
Overseas power cable (domestic power cable is included with the unit) For North America (PE3530083) For Europe (PE3530085) For shipments to China (CCC), please contact us.
|