Capacitive Non-contact Displacement Sensor
VT-5200シリーズ

VT-5200 series displacement gauge + VE series displacement sensor
VT-5210
VT-5220
VE Series Displacement Sensor
VE Series Displacement Sensor
VE Series Displacement Sensor
VE Series Displacement Sensor

The newly developed VE series capacitive gap detector combined with the VT series converter is a capacitive micro-displacement gauge that measures the distance and displacement to the measurement target (conductors such as iron, aluminum, copper, and various metal platings) with high precision and non-contact. It measures the gap and displacement between the sensor electrode's end face of the VE series sensor, installed opposite the target surface, and the capacitance formed by them. Two models are available: the VT-5210 with a response frequency of 4kHz and the VT-5220 with a high-speed response of 10kHz. These devices excel in measuring and controlling shaft vibrations and surface runout of rotating machines such as turbines, electric motors, and compressors, as well as the warpage, flatness, and thickness of conductors like silicon wafers and hard disks. Due to non-contact detection, they do not impose any load on the rotating or vibrating measurement targets, nor do they alter the vibration mode. The VE sensors are available in nine types for measurement ranges from 20μm to 8mm. The capacitive non-contact displacement gauge VT-5700 series (compact type for embedded use) → *The capacitive non-contact displacement gauges VT-5210 and VT-5220 have been discontinued.

Notice of Discontinuation of Sales

The VT-5210 and VT-5220 capacitive non-contact displacement gauges have been discontinued.

Features

  • Measurement range from 20μm to a maximum of 8 mm, wide range.

  • Measurement accuracy of 0.2% of full scale (F.S.) with a combination of converter and sensor.

  • Max 10 kHz high-speed frequency response - high-speed response even to dynamic displacement changes (VT-5220)

  • It does not affect the measurement target for non-contact measurement.

  • All conductors are measurable - they are not affected by factors such as color, roughness, reflectivity, or external light.

Case Studies

Axis vibration measurement

  • Axis vibration measurement

Surface vibration measurement

  • Surface vibration measurement

Warping, Deflection, Flatness Measurement

  • Warping, Deflection, Flatness Measurement

Specifications

Model name VT-5210 Discontinued  VT-5220 discontinued
detection method capacitive
Output 0~5 V/0~100 % F.S.
linearity ±0.2 % F.S. / 10~100 % F.S. ±0.25 % F.S. / 10~100 % F.S.
response frequency DC ~4 kHz DC ~10 kHz
Instruction Department 0 to 100 % / 20 divisions LED
Operating temperature range 0 to +40 °C (Accuracy guarantee range: 23 ± 2 °C)
Use power AC 100~240 V 10VA
Outline Dimensions 95 (W)×150(H) ×195(D) mm
Outline Dimensions
Weight approximately 2 kg
Price    

* F.S. (Full Scale): This is the maximum measurement range of the VE sensor.

VE Series Capacitive Gap Detector

standard type

Model name VE-5010 VE-1020 VE-1520 VE-3020 VE-8020
Measurement range (μm) ※1 50~500 100~1000 150~1500 300~3000 800~8000
Dimensions (mm) ※2 ⌀6 VE-5010 ⌀8 VE-1020 ⌀10 VE-1520 ⌀20 VE-3020 ⌀40 VE-8020(Retaining part ⌀20)
cable length 1.5m (cable integrated type) Separate sale VL-1520/VL-1521 cable (1.5m)
temperature coefficient k1=1.7×10-5、k2=3.4×10-5
Operating temperature range ※3 0~+80 ℃
Price          

Common clamp shape type, ⌀10mm

Model name VE-2011 VE-5011 VE-1021 VE-8021
Measurement range (μm) ※1 20~ 200 50~500 100~1000 800~8000
External dimensions of the measuring section (mm) ※2 ⌀3
(retention part ⌀10)VE-2011
⌀6
(retaining part ⌀10)VE-5011
⌀8
(retention part ⌀10)VE-1021
⌀40
(holding part ⌀10)VE-8021
cable length Separate sale VL-1520/VL-1521 cable (1.5m)
temperature coefficient k1=1.7×10-5、k2=3.4×10-5
Operating temperature range ※3 0~+80 ℃
Price        

*1 The measurement range indicates the maximum gap from the sensor's end face to the measurement target.
*2 The measurement surface of the target object requires an area larger than the sensor's outer diameter.
*3 The operating temperature range refers to the temperature range in which the sensor can function, but it is not the temperature range where linearity is guaranteed. To ensure linearity, please use at a temperature of 23±2°C.

signal cable sold separately

VL-1520(両端ストレートタイプコネクター)
VL-1520 (both ends straight type connector)
VL-1521(L型コネクター-ストレートタイプコネクター)
VL-1521 (L-shaped connector - straight type connector)