E-Mail: overseas@onosokki.co.jp

Select your region & language

Global

Region

CF-5200 Frequently Asked Questions

Question number FAQ-0504

The reading from the sound level meter is different from the overall value of the CF (Call Field).

The cause of the difference is

(1) Sound level meter side
Changing the measurement level range between CAL and measurement will result in a dB difference. (This information will not be received by the CF.)

(2) When the frequency range is 20 kHz or lower, sound components above the frequency range are cut off, resulting in a difference due to the amount of the cut.

(3) The difference arises because the time constants of the sound level meter (FAST, SLOW) and the time constant of the CF are different.

The CF time constant corresponds to the measurement time for the data length (2048 points).

Related instruction manual chapters

3.5 Anti-aliasing filter
3.6 Setting the Analysis Data Length and Resolution

Last updated: 2000-04-10