E-Mail: overseas@onosokki.co.jp

Select your region & language

Global

Region

Digital Gauges: Frequently Asked Questions

Question number FAQ-0120

Changing the measurement force of a linear gauge sensor

At our company, the measuring force is defined as the force exerted by the measuring probe of a linear gauge sensor against the object being measured. This measuring force is generated by a spring built into the linear gauge sensor, and the return force when the spindle is fully depressed is used as the measuring force. The unit is expressed in Newtons (N), and the measurement method conforms to JIS B 7503 Dial Gauge. Please refer to this link for the definition of measuring force.

 

The measuring force can be changed by replacing the built-in spring, which is useful when you want to reduce the measuring force for a deformable object, or when you want to increase the measuring force to make the spindle return more easily when using a linear gauge sensor mounted horizontally or upwards.
Please refer to the table below for the adjustable range of the measuring force (excluding the BS-1210/1310 and GS-7710 series).

 

If you wish to change the measuring force, please request this when purchasing the linear gauge sensor.

When placing an order,


1. At the beginning of pushing (around 1 mm), ○○N

2. At the maximum indentation position (near the maximum measurement amount), △△N

3. Ensure it returns completely when used in an upward position.

Therefore, please specify the change in measuring force using one of the following options: 1 to 3.
Standard modifications cost ¥15,000 (excluding tax) and take approximately 10 days to complete.
For further details, please contact your nearest branch office.
Depending on the modifications, the mounting method may differ for horizontal and upward orientations.
Please note that the spindle may not return completely.

 

 

Model name standard measuring force
(Catalog value)
The value is when it is downward.
Adjustable range of measuring force
(The value in parentheses indicates the measuring force near the entrance.)
The value is when it is downward.
BS-1210
BS-1310
1.47 N or less Not possible
*1
GS-7710A
GS-7710NA
1.08 N or less Not possible
GS-1713A
GS-1813A
GS-6713A
GS-6813A
1.3 N or less 0.7 ~ 2.2 N
(0.3 ~ 1.0 N)
*1
GS-1730A
GS-1830A
GS-6730A
GS-6830A
1.9 N or less 1.0 ~ 3.1 N
(0.3 ~ 0.9 N)
*1
GS-4713A
GS-4813A
1.8 N or less 0.8 ~ 2.5 N
(0.4 ~ 1.1 N)
GS-4730A
GS-4830A
2.4 N or less 1.5 ~ 3.9 N
(0.5 ~ 0.9 N)
GS-3813B 2.3 N or less 0.8 ~ 3.0 N
(0.3 ~ 1.4 N)
GS-3830B 2.7 N or less 1.5 to 2.5N
(0.4 to 0.8 N)
Upward modification is not possible.
GS-5050A
GS-5051A
2.9 N or less 1.8 ~ 6.2 N
(1.3 ~ 3.2N)
GS-5100A
GS-5101A
5.2 N or less 3.1 ~ 7.2 N
(1.6 ~ 3.4 N)

(Variation: ±0.15 N)

Note *1
For constant pressure modifications, a separate consultation will be required, so please contact your nearest branch office.
Constant pressure modification is the process of making the measuring force, which normally varies depending on the spindle position, constant throughout the entire stroke (within ±0.1 N, without dustproof rubber).

Last updated: 2019-03-27