Vibration Comparator Frequently Asked Questions
Question number FAQ-1394
Setup Procedure Overview
This explains the setup procedure for first-time users.
While you can also access this function using the function buttons on the front panel, I will explain the procedure for setting it through the menu.
● Outline of the setup procedure
Connect Accelerometer and various signal lines to the rear terminals. The VC-2100 does not have a power switch. Please note that the power will be turned on as soon as the power is connected.
Set the sensor sensitivity to, for example, 5.00mV/ms^-2, using the test sheet included with Accelerometer.
Set the measurement mode to PEAK or PEAK HOLD, and adjust the bar indicator so that it does not go OVER, but stays within a range of approximately 50-60%.
Set the bandpass frequency so that the necessary signal can be detected.
To make it easier to read the vibration values, set the measurement mode.
The appropriate judgment level is set based on the measured vibration value.
Please refer to the PDF simplified operation manual for details.