Skip to main content

Select your region & language

Global

Region

Vibration Comparator Frequently Asked Questions

Question number FAQ-1391

Does the DC output voltage value change depending on whether the measurement mode is PEAK or RMS?

It doesn't change. The DC output is always output according to the magnitude of the RMS value.

The input signal is a sine wave with an amplitude of 0.349 km/s 2 In that case, the effective value is 0.247 km/s 2 is
Full scale is 1,000 km/s 2 This would result in a voltage value of 1.24 V.

Last updated: 2017-10-20