Accelerometer Frequently Asked Questions
Question number FAQ-1386
When integrating a signal with the PS-1300, what is calibration method using the DS-0221 general-purpose FFT software?
As described on page 20 of the PS-1300 instruction manual, calculate the displacement value per 5V.
When the sensitivity of the sensor used is [0.9mV/(m/s²)] and the gain setting of the PS-1300 is [10];
■ Measurable maximum acceleration, maximum velocity, maximum displacement
(When using a sensor with a sensitivity of 1 mV/(m/ s²))
In the case of acceleration (ACC m/ s²)
| Gain | 1 | 2 | 5 | 10 | 20 | 50 | 100 |
|---|---|---|---|---|---|---|---|
| Maximum acceleration (m/s 2) | 5000 | 2500 | 1000 | 500 | 250 | 100 | 50 |
- In the case of speed (VEL mm/s)
| Gain | 1 | 2 | 5 | 10 | 20 | 50 | 100 |
|---|---|---|---|---|---|---|---|
| Maximum speed (mm/s) | 5000 | 2500 | 1000 | 500 | 250 | 100 | 50 |
- In the case of displacement (DISP mm)
| Gain | 0.1 | 0.2 | 0.5 | 1 | 2 | 5 | 10 |
|---|---|---|---|---|---|---|---|
| Maximum displacement (mm) | 500 | 250 | 100 | 50 | 25 | 10 | 5 |
For sensors with a sensitivity other than 1 mV/(m/ s²), please check the following formula before use.

From the above formula:
V/EU setting value = Sensor sensitivity × 5 ÷ Maximum displacement value determined by the gain setting (see table above)
(Sensor sensitivity is expressed in mV/(m/ s²) units)
In the example above, the actual settings are:
In the DS-0221's input menu, under "Units/ calibration," set the unit name to "mm" and the physical quantity to "0.9V/EU." The same applies to speed, where the unit name is mm/s.
Note that this calibration method is the same for other FFT analyzers, such as the CF-7200.
Last updated: 2007-08-20