Acceleration Pickup Frequently Asked Questions
Question number FAQ-1386
The calibration method using the DS-0221 general-purpose FFT software when integrating signals with the PS-1300 is.
As stated on page 20 of the PS-1300 manual, calculate the displacement value per 5V.
The sensitivity of the used sensor is [0.9mV/(m/s 2] when the gain setting of the PS-1300 is [10];
■Maximum measurable acceleration, maximum speed, maximum displacement
(When using a sensor with a sensitivity of 1 mV/(m/s 2))
In the case of acceleration (ACC m/s 2)
| Gain | 1 | 2 | 5 | 10 | 20 | 50 | 100 |
|---|---|---|---|---|---|---|---|
| Maximum acceleration (m/s 2) | 5000 | 2500 | 1000 | 500 | 250 | 100 | 50 |
- In the case of speed (VEL mm/s)
| Gain | 1 | 2 | 5 | 10 | 20 | 50 | 100 |
|---|---|---|---|---|---|---|---|
| Maximum speed (mm/s) | 5000 | 2500 | 1000 | 500 | 250 | 100 | 50 |
- In the case of displacement (DISP mm)
| Gain | 0.1 | 0.2 | 0.5 | 1 | 2 | 5 | 10 |
|---|---|---|---|---|---|---|---|
| Maximum Displacement (mm) | 500 | 250 | 100 | 50 | 25 | 10 | 5 |
For sensors other than those with a sensitivity of 1 mV/(m/s 2), please confirm using the following formula before use.

From the above expression;
V/EU setting value = Sensor sensitivity × 5 ÷ Maximum displacement value determined from gain settings in the table above.
(Sensor sensitivity is a value in mV/(m/s 2))
In the above example, the actual setting is;
In the DS-0221 input menu, set the unit name to "mm" and the physical quantity to "0.9V/EU" in the "Unit calibration" input menu of the DS-0221. The same applies to speed, where the unit name is mm/s.
Furthermore, this calibration method is the same for other FFT analyzers, such as the CF-7200.