Accelerometer Frequently Asked Questions

Question number FAQ-1386

When integrating a signal with the PS-1300, what is calibration method using the DS-0221 general-purpose FFT software?

As described on page 20 of the PS-1300 instruction manual, calculate the displacement value per 5V.

When the sensitivity of the sensor used is [0.9mV/(m/s²)] and the gain setting of the PS-1300 is [10];


■ Measurable maximum acceleration, maximum velocity, maximum displacement
(When using a sensor with a sensitivity of 1 mV/(m/ ))

  • In the case of acceleration (ACC m/ )

Gain125102050100
Maximum acceleration (m/s 2)50002500100050025010050
  • In the case of speed (VEL mm/s)
Gain 1 2 5 10 20 50 100
Maximum speed (mm/s) 5000 2500 1000 500 250 100 50
  • In the case of displacement (DISP mm)
Gain 0.1 0.2 0.5 1 2 5 10
Maximum displacement (mm) 500 250 100 50 25 10 5

For sensors with a sensitivity other than 1 mV/(m/ ), please check the following formula before use.

From the above formula:

V/EU setting value = Sensor sensitivity × 5 ÷ Maximum displacement value determined by the gain setting (see table above)

(Sensor sensitivity is expressed in mV/(m/ ) units)

In the example above, the actual settings are:

In the DS-0221's input menu, under "Units/ calibration," set the unit name to "mm" and the physical quantity to "0.9V/EU." The same applies to speed, where the unit name is mm/s.

Note that this calibration method is the same for other FFT analyzers, such as the CF-7200.