Torque Meter TS-2100/2600 Frequently Asked Questions
Question number FAQ-1317
No analog voltage output.
| cause | Confirmation items | Processing method |
|---|---|---|
| A negative voltage was output due to a setting error in the torque detector. | Is the torque display negative, or is the output a negative voltage? | See symptom "Negative torque is displayed". |
| Inappropriate connection load | Is the connected load 10kΩ or more? | Change connection load |
Last updated: 2026-07-15