Question number FAQ-1317

No analog voltage output.

causeConfirmation itemsProcessing method
A negative voltage was output due to a setting error in the torque detector.Is the torque display negative, or is the output a negative voltage?See symptom "Negative torque is displayed".
Inappropriate connection loadIs the connected load 10kΩ or more?Change connection load