Question number FAQ-1308

To connect the TS-2600 BCD output to the sequencer.

Below is an example of wiring.

■BCDout specifications

suitable plugDX40-50P (Hirose Electric)
same plug coverDX-50-CV1 (Hirose Electric Co., Ltd.)
We have the AA8012 model BCD cable, 5m, one side open, for ¥20,000.

output formatOpen Collector Positive Logic
Signals marked with an asterisk (*) are of negative logic.

IC:74LS07Voltage resistance: 24V max,
Sink current: 32mA max,
Residual voltage: 0.6V
PIN numbersignal namePIN numbersignal name

1
2
3
4

5
6
7
8
Data output (torque section)
1×10 0
2×10 0
4×10 0
8×10 0

1×10 1
2×10 1
4×10 1
8×10 1

26
27
28

29
30
31
32

33
Data output (rotating part)
2×10 1
4×10 1
8×10 1

1×10 2
2×10 2
4×10 2
8×10 2

1×10 3
9
10
11
12

13
14
15
16
1×102
2×102
4×102
8×102

1×103
2×103
4×103
8×103
34
35
36

37
38
39
40
41
2×103
4×103
8×103

NC
NC
NC
NC
NC
17
18
19
20

21
22
23
24

25
NC
NC
NC
NC
Data output (rotating part)
1×10 0
2×10 0
4×10 0
8×10 0

1×10 1
42
43
44
45

46
47
48
49
50
Torque polarity output "−"
Torque polarity output "+"
NC
NC

NC
Hold input *
Busy input *
Print command output *
Common ( 0V )

Hold Input *
While entering the Lo level, the BCD output data is held.

Busy Input *
Used when connecting to the printer, inputting a Lo level signal during printing.

Print command output *
When receiving a hold signal or busy signal, it outputs a negative logic signal with a pulse width of approximately 100 to 150 μs.