Torque Meter Frequently Asked Questions
Question number FAQ-1308
To connect the TS-2600 BCD output to the sequencer.
Below is an example of wiring.

■BCDout specifications
| suitable plug | DX40-50P (Hirose Electric) |
| same plug cover | DX-50-CV1 (Hirose Electric Co., Ltd.) |
| We have the AA8012 model BCD cable, 5m, one side open, for ¥20,000. |
| output format | Open Collector Positive Logic |
| Signals marked with an asterisk (*) are of negative logic. | |
| IC:74LS07 | Voltage resistance: 24V max, Sink current: 32mA max, Residual voltage: 0.6V |
| PIN number | signal name | PIN number | signal name |
|---|---|---|---|
| 1 2 3 4 5 6 7 8 | Data output (torque section) 1×10 0 2×10 0 4×10 0 8×10 0 1×10 1 2×10 1 4×10 1 8×10 1 | 26 27 28 29 30 31 32 33 | Data output (rotating part) 2×10 1 4×10 1 8×10 1 1×10 2 2×10 2 4×10 2 8×10 2 1×10 3 |
| 9 10 11 12 13 14 15 16 | 1×102 2×102 4×102 8×102 1×103 2×103 4×103 8×103 | 34 35 36 37 38 39 40 41 | 2×103 4×103 8×103 NC NC NC NC NC |
| 17 18 19 20 21 22 23 24 25 | NC NC NC NC Data output (rotating part) 1×10 0 2×10 0 4×10 0 8×10 0 1×10 1 | 42 43 44 45 46 47 48 49 50 | Torque polarity output "−" Torque polarity output "+" NC NC NC Hold input * Busy input * Print command output * Common ( 0V ) |
Hold Input *
While entering the Lo level, the BCD output data is held.
Busy Input *
Used when connecting to the printer, inputting a Lo level signal during printing.
Print command output *
When receiving a hold signal or busy signal, it outputs a negative logic signal with a pulse width of approximately 100 to 150 μs.
Last updated: 2000-06-09