Torque Meter Frequently Asked Questions
Question number FAQ-1268
Wiring work and noise
When noise is superimposed on the torque signal, for example;
- The display shows significant fluctuations even when there is no load.
- Despite a constant rotation speed and constant load torque, the torque display and analog output voltage fluctuate significantly.
- When the power to the drive motor and load device is turned off, the display stabilizes at zero, but when the power is turned on, the fluctuations become larger.
These symptoms may manifest as noise. It is necessary to take preventative measures against such noise during installation and wiring. The following is an excerpt from the "Precautions for Torque Meter Wiring Installation" section of the TS-3200 display unit instruction manual. Please refer to it.
In recent years, there has been an increase in test equipment that uses or measures inverter motors. Due to their operating principle, inverters generate noise, which can affect adjacent equipment, and torque meters are no exception. The degree of impact is not uniform and depends on factors such as the amount of noise generated by the inverter, the condition of the signal lines, the installation conditions, and the distance of the equipment from the inverter. However, problems caused by inverter noise can be reduced by referring to the following precautions regarding wiring work in advance. We also ask that you refer to the following when performing construction work even in cases other than inverter noise.
|