Question number FAQ-1228

What is the delay time of the AC output on the LA-3000 series?

It takes about 5 ms.
If you need a signal with no delay, use the Through output.

Delay time in Single mode

5.0 ms (input - AC-Z OUT)
5.0 ms (input - AC OUT)

Delay time when using a 1/3 octave filter

Center frequency (Hz)Sampling frequency (Hz)Input-ACOUT delay (ms)
4 kHz64 kHz5.1 ms
2 kHz32 kHz5.6 ms
1 kHz16 kHz6.4 ms
500 Hz8 kHz7.9 ms
250 Hz4 kHz11.2 ms
125 Hz2 kHz17.9 ms
63 Hz1 kHz30.8 ms
31.5 Hz500 Hz56.3 ms

Delay time when using a 1/1 octave filter

Center frequency (Hz)Sampling frequency (Hz)Input-ACOUT delay (ms)
4 kHz64 kHz5.1 ms
1 kHz32 kHz5.4 ms
500 Hz16 kHz6.1 ms
250 Hz8 kHz7.3 ms
125 Hz4 kHz9.7 ms
63 Hz2 kHz15.2 ms
31.5 Hz1 kHz25.0 ms