Skip to main content

Select your region & language

Global

Region

sound level meter Frequently Asked Questions

Question number FAQ-1228

What is the delay time of the AC output for the LA-3000 series?

It is about 5 ms.
If you need a signal without delay, please use the Through output.

Delay time in single mode

5.0 ms (Input - AC-Z OUT)
5.0 ms (Input - AC OUT)

Delay time of 1/3 octave filter

Center frequency (Hz)Sampling frequency (Hz)Input Delay (ms)
4 kHz64 kHz5.1 ms
2 kHz32 kHz5.6 ms
1 kHz16 kHz6.4 ms
500 Hz8 kHz7.9 ms
250 Hz4 kHz11.2 ms
125 Hz2 kHz17.9 ms
63 Hz1 kHz30.8 ms
31.5 Hz500 Hz56.3 ms

Delay time during 1/1 octave filter.

Center frequency (Hz)Sampling frequency (Hz)Input - ACOUT delay (ms)
4 kHz64 kHz5.1 ms
1 kHz32 kHz5.4 ms
500 Hz16 kHz6.1 ms
250 Hz8 kHz7.3 ms
125 Hz4 kHz9.7 ms
63 Hz2 kHz15.2 ms
31.5 Hz1 kHz25.0 ms

Last updated date: 2014-12-03