The correction values for A-weighting are listed below. These correction values are specified in "JIS C1509". Please add the following values to the flat value (dB).

Frequency (Hz)Correction value (dB)
20-50.5
25 -44.7
31.5-39.4
40 -34.6
50-30.2
63 -26.2
80-22.5
100 -19.1
125-16.1
160 -13.4
200-10.9
250 -8.6
315-6.6
400 -4.8
500-3.2
630 -1.9
800-0.8
1000 0.0
12500.6
1600 1.0
20001.2
2500 1.3
31501.2
4000 1.0
50000.5
6300 -0.1
8000-1.1
10000 -2.5
12500-4.3
16000 -6.6
20000-9.3