Tachometer Frequently Asked Questions

Question number FAQ-0939

TM-3140 "When you want to turn off (open) the relay in case of an abnormality."

Set​ ​CMP1 to UPPER and CMP2 to OK.


When the displayed value is less than UPPER
CMP1 will have the relay OFF, and CMP2 will have the relay ON.


When the displayed value is greater than UPPER
CMP1 will turn the relay ON, and CMP2 will turn the relay OFF.

This can be used when you want to turn off a relay at maximum rotation speed, for example.

■ UPPER setting device
When UPPER ≤ the displayed value, the relay turns ON.

■ LOWER setting device
When LOWER ≤ the displayed value, the relay turns ON.

■ OK setting device
When UPPER and LOWER are OFF, the relay turns ON.
When either one is ON, the relay is OFF.