Frequently Asked Questions about Magneto-Electric Rotation Detector
Question number FAQ-0848
Method for securing the MP-981
Method for securing the MP-981
(1) Mounting bracket
Use a non-magnetic material and secure it so that it does not move due to vibration or other factors.
Do not thread the mounting holes; instead, use the nuts provided with the MP-981 to secure it.
A nut tightening force of 15-20 Nm is recommended.
(2) Alignment
Align the alignment mark (▲) on the MP-981 so that it is approximately in the center of the gear teeth.
Please install the gear shaft and the detector so that they are at a right angle.
The distance from the gear should be approximately 1 mm.



Last updated: 2002-02-15