FV Converter Frequently Asked Questions

Question number FAQ-0842

Why is the output signal of the FV-1500 not dropping to 0V?

The response time of the FV-1500's output signal is the input frequency cycle time + 3.5 μs.
The input frequency range is 0.2 Hz to 320 kHz, and when the input signal is 0.2 Hz, the output update time is approximately 5 seconds.
If there is no input signal, the voltage and current values of the output signal are retained.
(If no stop time is set for the FV-1500) The voltage will return to 0V 5.5 seconds after the latest output update.
(When current is output, the values will be 0 mA and 4 mA.)