Frequently Asked Questions about FV Converters
Question number FAQ-0836
The relationship between the frequency range and voltage output of the FV-1100 F/V converter is
The standard specification of the FV-1100 is max 10kHz.
When the operating frequency is below 50% of the frequency range, the output voltage decreases while the ripple * increases, resulting in poor accuracy within the usable range.
It is recommended to modify the system so that the frequently used frequency range approaches the full-scale frequency.
| (for example) | In the 10kHz range, the accuracy is ±0.2%, which means If the operating range is below 1kHz, When modifying 1kHz to full scale |
The usage range of the FV converter should be considered approximately between 10% and 100% of the full-scale frequency range.
Regarding Ripple, please refer to the following PDF file.
Last updated date: 2002-02-15