Question number FAQ-0832

While measuring fuel flow with the FX series flow detector, an error message "LEVEL ERROR" was displayed on the FM series flow meter. Why is that?

"LEVEL ERROR" lights up when the fuel level in the FX series detection unit reaches the lower limit (indicating fuel depletion) or the upper limit (indicating overflow). This occurs when the fuel consumption of the test specimen is high, leading to insufficient fuel supply to the FX, or when the consumption flow rate of the test specimen is low, resulting in excessive supply to the FX. Please check if fuel is being supplied properly.

If 'LEVEL ERROR' is constantly displayed, there may be fuel accumulated in the reference tube. Please refer to the manual, open the cover, and try draining the fuel from the reference tube.