When checking the user manual for each model of the FP series, there are SIG1 and SIG2 connections. Is it acceptable to connect only SIG1 if there is no need to detect backflow of the measured fluid?
There is a problem.
Since SIG1 is output in both forward and reverse rotation, it cannot be measured correctly with only the connection of SIG1.
To measure correctly, a circuit that processes both SIG1 and SIG2 is required.
Since SIG1 is output in both forward and reverse rotation, it cannot be measured correctly with only the connection of SIG1.
To measure correctly, a circuit that processes both SIG1 and SIG2 is required.
Last updated: 2020-02-17