Basic FAQ for FFT Analyzer
Question number FAQ-0774
I want to evaluate in dBm. What are the settings for the calibration value?
The absolute level value of power is defined as 0 dBm for a value of 1 mW.
The V/EU value to be set will be as follows, depending on the impedance of the circuit being measured.
| Impedance Z (Ω) | V/EU |
|---|---|
| 50 | 0.224 |
| 75 | 0.274 |
| 600 | 0.775 |
Last updated: 2014-04-14