DS-3000 Frequently Asked Questions
Question number FAQ-0738
How can I apply frequency correction as defined by "ISO 5349-1 (Measurement Standard for Hand-Arm Vibrations)" using the DS-0323 Real-Time Octave Software?
First, please download the following CSV file.
This is a definition file for user filters used to correct hand-arm vibrations.
■Operating Procedure
- From the menu, select [Analysis Settings] → [Frequency Weighting].
Select "User Defined" for frequency weighting.
Open the user-defined filter settings "Open" and load the downloaded CSV file. - Set the octave band width to "1/3".
The setup is now complete.
For detailed instructions, please refer to the operating manual provided below.
The PDF procedure manual is available here (Hand-arm vibration measurement procedure compliant with JIS B 7761-3 (ISO 5349-1)).