Skip to main content

Select your region & language

Global

Region

DS-3000 Frequently Asked Questions

Question number FAQ-0733

How can I change the output voltage of the DS-0321 time record signal output?

Select "Full Scale of A/D Data" as the amplitude mode for the time record signal output, and then set the amplitude value.

• Amplitude during recording: The output will retain the voltage magnitude of the recorded signal.

- A/D Data Full Scale: Outputs the voltage range value (A/D full scale) recorded at any desired voltage. The maximum configurable voltage is 10 V.

For example, if the recording range was 0.1 Vrms (0.141 V), and you set the amplitude [V] to 5,
5 ÷ 0.141 = 35.5 times the voltage will be output.

Last updated: 2019-07-09