E-Mail: overseas@onosokki.co.jp

Select your region & language

Global

Region

DS-3000 Frequently Asked Questions

Question number FAQ-0729

Please tell me about the necessary equipment and precautions for FRAME LINK2 (connecting 4 chassis).

In addition to the PC and the DS-3200 enclosure, the following equipment is required:
* The DS-3100 (older model) is not supported.

 

Required hardware

DS-0392A inter-chassis connection interface (See diagram below for verification method)

DS-0394 inter-chassis connection box

 

 

Required software

DS-0321A FFT analysis (See diagram below for verification method)

 

 

 

 

Required cables

-AX-9035 or AX-9036 inter-chassis connection interface cable (0.75 m or 2 m)

AX-9041 USB cable (2m)

 

Recommended USB hub

DS-0393 Inter-chassis USB hub

(It is also possible to connect it directly to the PC without using a USB hub.)

 

FRME LINK2 Configuration Example

 

 

 

Important points regarding FRAME LINK2

  • This product is only compatible with the DS-3200 (current model). The DS-3100 (older model) is not compatible.
  • The inter-chassis connection box (DS-0394) is absolutely necessary.
  • The DS-3200 (main unit) needs to be equipped with an inter-chassis connection interface (DS-0392A).
  • Commercially available USB hubs cannot be used. You must use the enclosure-to-enclosure USB hub (DS-0393).
  • A license for the DS-0321A (FFT analysis function) software is required. DS-0321 will not operate on its own. ●Operates in FFT-A analysis mode. It will not operate in any other mode.
  • ●It is also possible to connect multiple enclosures and a PC directly via USB without using the inter-enclosure USB hub (DS-0393).
  • It is not possible to use inter-chassis interface cables of different lengths together.

 

Other:How to check if FRAME LINK2 (4-chassis connection) is not working.