DS-2000 Frequently Asked Questions
Question number FAQ-0717
What voltage does the monitor output on the back of the DS-2000 series unit output? Also, is there a filter applied?
The output voltage will be 1.414V (1 Vrms) when the voltage level is the same as the full scale of the voltage range. Additionally, the A-weight, B-weight, and C-weight analog filters set in Input → Voltage Range Setting → Filter Selection will be applied.
for example;
| Voltage range: A Vrms (RMS) Input signal: BV (single amplitude) Monitor output voltage: CV (single amplitude) Therefore, it can be calculated using proportional calculation as follows: (example) C=1÷3.16=0.316 V (half amplitude) It will be. |
Note
The voltage range can be set to display in Vrms (RMS), dBVrms (logarithmic), or V (single amplitude). The above formula is for Vrms (RMS). If you are using dBVrms (logarithmic) or V (single amplitude), please convert it to Vrms (RMS).
■Filter selection screen

Last updated: 2006-08-21