DS-2000 Frequently Asked Questions

Question number FAQ-0690

How can I disable calibration when performing offline analysis on a DS-0221 after recording an ORF file with unit calibration using the DS-0250 throughput function?

Set the unit calibration of the DS-0221 to "calibration: ON" and set "Physical Value: 1".

supplementary explanation

  • When unit calibration is turned on in DS-0221, calibration value of DS-0221 takes precedence over calibration value in the DS-0250 record.
  • Turning off the unit calibration for DS-0221 will revert to calibration used for DS-0250.

For example, setting the physical value of the DS-0221's unit calibration to "1 EU/V" (or 1V/EU) will result in the same value as the input voltage of the DS-0250. Setting it to "5 EU/V" will activate this calibration value. In this way, the DS-0221's unit calibration can be arbitrarily reset during offline analysis.

Last updated: 2006-11-20