DS-2000 Frequently Asked Questions

Question number FAQ-0681

To apply frequency correction as specified in "ISO 2631-1 (Measurement standard for whole-body vibration)" and "ISO 5349-1 (Measurement standard for hand-arm vibration)" using DS-0223 real-time octave software.

First, please download the following zip file.

When you unzip the downloaded file, three files will be created.

  • ISO2631-Wd.fit
  • ISO2631-Wk.fit
  • hand-ISO5349.fit

"ISO2631-Wd.fit" is the definition file for user filters for whole-body vibration in the horizontal direction, "ISO2631-Wk.fit" is for whole-body vibration in the vertical direction, and "hand-ISO5349.fit" is for the correction characteristics of hand-arm vibration.

Each file is in text format, so you can open it with software like Notepad to check the values.

Operation Procedure

Select [Input] → [User Filter Settings] from the menu.

Set the octave to "1/3" and click the "Load" button. Please load the definition file you downloaded earlier.

The settings are complete.

Last updated date: 2007-07-17