Question number FAQ-0680

The DS-0222 tracking analysis software cannot perform rotation and fixed-time tracking, and the DS-0223 real-time octave software cannot perform time trend and tracking.

phenomenon

  • When trying to perform tracking (time-based tracking) with DS-0221, pressing the 'START' button causes it to stop immediately. The number of store blocks remains at 0.
  • When trying to perform time trend and rotation tracking with DS-0223, the process stops before reaching 0% or 100% even after pressing the 'START' button. The trend graph is displayed, but the data is chaotic.

Reason:

A large number of temporary files (DS21*.tmp) created by the offline analysis library remain, preventing the creation of new temporary files.


Countermeasures:

  1. If there is a file named DS21*.tmp in the temporary folder, please delete it. The default path for the temporary folder is as follows: <UserName> will be replaced with the name of the logged-in user.
    • For Windows XP: C:\Documents and Settings\<username>\Local Settings\Temp
    • For Windows 7: C:\Users\<username>\AppData\Local\Temp
  2. If you are worried about directly manipulating the temporary folder, please use the following tool to clean up "temporary files."
    [All Programs]→[Programs]→[Accessories]→[System Tools]→[Disk Cleanup]

Last updated: 2010-03-11