DS-2000 Frequently Asked Questions

Question number FAQ-0669

Is it possible to read CF series record memory files with the DS-0221 and perform offline analysis? How can this be done?

The DS-2000 series offline analysis software can read time record memory data taken with previous CF series cameras.

■Compatible Models

This is possible with the following CF series analyzer models that can save in DOS format.

CF-6400, CF-5200, CF-3200/3400, DS-9100

■Method

  • Let's name the filenames of the record memory obtained using CF cards Aa001.rcd and Aa002.rcd.
    (In the case of 2ch data)
  • Create a file named Aa001.rci in the same folder as the rcd file.
    (The file name can be anything you like.)
  • As the contents of the Aa001.rci file, use a text editor (such as Notepad):

    Aa001.rcd
    Aa002.rcd

    Write the following two lines. (The rci file is an index file for referencing the rcd file.)

  • Start up DS-0221;

    File menu → "Open offline analysis data"

    Then, set the file type to DS RecordIndex File (*.rci) and open it by specifying the above Aa001.rci file.

Last updated: 2006-05-22