Frequently Asked Questions for CF-9200/CF-9400
Question number FAQ-0652
An 'Error Change Folder' message appears every time the power is turned on.
phenomenon
Every time the power is turned on for the CF-9200 and CF-9400,
Error Change Folder(Panel):E:Onosokki CF-9000CF-9000Panel
A warning message is displayed.
The “Panel” portion at the end of the message
“Panel (-1)”
“DataMem1” and “DataMem1 (-1)”
“AutoSeq” and “AutoSeq (-1)”
“Record” and “Record (-1)” may also be displayed.
Avoidance Procedure
To delete a message, please switch to the configuration mode (CF-9000 Configuration Mode) first. Follow the steps to operate the CF-9200 and CF-9400. Additionally, to prevent power disconnection during the operation, connect the AC adapter that comes with the CF-9200 and CF-9400 and perform the work while supplying power from the AC adapter.
Transition to configuration mode.
Tap the soft key from the top mode (HOME) in the order of [System] > [Env Set] > [Config Mode], and then tap OK on the message dialog box that appears.


After the CF-9200 and CF-9400 systems automatically shut down, the CF-9000 Configuration Mode screen is displayed.
Exit configuration mode
Tap 'Exit' on the CF-9000 Configuration Mode screen to exit the CF-9000 Configuration Mode.

CF-9000 Configuration Mode screen
Tapping the OK button on the confirmation dialog box that appears next will automatically shut down the CF-9200 and CF-9400.

Confirmation
Press the POWER switch of the CF-9200 and CF-9400 to turn on the power, and confirm that no warning messages are displayed.
*Even if this warning message is displayed, it will not affect the functionality, performance, or measurement analysis accuracy of the FFT analyzer. There are no issues with the data recorded so far. Additionally, measurements will not be affected even if this workaround is not implemented.
Last updated date: 2015-11-18