CF-7200 Frequently Asked Questions

Question number FAQ-0639

The error message "Warning (114): no sensor" appears and measurement is not possible.

Phenomenon

When a CCLD (Constant Current Line Drive 4mA, DC24V) compatible Accelerometer is connected, the input is set to CCLD, and the "START" button is pressed, the CCLD setting is canceled and the error message "Warning (114): no sensor" is displayed.

■ Causes and countermeasures

If the CF-7200 cannot communicate with Accelerometer for any reason, its disconnection detection function will display the error message shown above. Possible causes include "signal cable not connected," "signal cable disconnected," or "Accelerometer malfunction." Try swapping the wiring with a working one, or replacing it with a different one and try again.

Last updated: 2009-06-15