Can trigger functions be used on record data?

You can use the trigger function on record data. However, the trigger source must be either CH1 or CH2. Additionally, the recording time for the record data must be at least three times the duration of one frame during analysis; if it is shorter, the trigger will not activate.

For example, in the case of a frequency range of 100 Hz with 2048 sample points;

One frame is calculated as 2048 ÷ (100 × 2.56) = 8 seconds, and if this is tripled, it results in 8 × 3, meaning at least 24 seconds of recording data time is required.

Last updated: 2007-08-20