Question number FAQ-0606

The PP value of the time-domain waveform and the overall PP value of the order spectrum do not match.

The peak of a time waveform is a combination of various frequency components.

The overall value obtained from the FFT does not fundamentally match the peak value of the time waveform.

Last updated: 2000-04-10