E-Mail: overseas@onosokki.co.jp

Select your region & language

Global

Region

CF-5200 Frequently Asked Questions

Question number FAQ-0603

(1) What happens when dBVr is converted to voltage? (2) What happens when the time-domain waveform is converted to amplitude reading in dB?

(1) dBVr=10LOG {Vrms/(reference)} 2 = 20LOG (Vrms)
The baseline is the RMS value of 1Vrms

Vrms=10dbVr/20

(2) Since the time waveform is a combination of various frequencies, caution is required when calculating dB from the waveform peak.

The calculation assumes PP = 1.39mV.

The amplitude (0-P) is 0.695mV, therefore, similar to (1)
dBV=20LOG (0.695× 10−3)
The standard is 1V per amplitude.

However, the frequency is unknown.

Related instruction manual chapters

4.3.1 Y-axis units of the spectrum

Last updated: 2000-04-10