Question number FAQ-0592

Phase correction cannot be performed at frequencies below 10kHz, but is it possible to correct at 50Hz?

Phase shift becomes severe on the high-frequency side, but at 50Hz, sufficient accuracy is achieved without phase correction.

Therefore, it is designed to function above the 10 kHz range.

Chapter of the relevant instruction manual

Phase Adjust