Skip to main content

Select your region & language

Global

Region

CF-5200 Frequently Asked Questions

Question number FAQ-0589

The inter-channel phase correction function does not correct even when the PHASE ADJ function is ON (remains at 0.26 degrees). A 50kHz square wave 1V is being input.

(1) Input 50kHz SIN wave and average more than 10 times, displaying FRF PHASE.
By adjusting PHASE, correction confirmation can be performed.

(2) The operation involves inputting random signals as calibration signals into chA and chB to measure the average frequency response function. In the POUSE state, turn on PHASE ADJUST.

(3) The phase correction accuracy at PHASE ADJUST ON is within the same voltage range
(typical value) ±0.1xxdeg at 20kHz or above
The value of xx is unknown

Chapter of the relevant instruction manual

4.3.4 Phase Adjust
Specification

Last updated date: 2006-01-16