CF-5200 Frequently Asked Questions

Question number FAQ-0585

I want to display the transfer function, but I want to perform the test using an impulse hammer.

The transfer function is measured using the frequency response function (FRF).

The following procedure is used to measure the transfer function (frequency response function) using an impulse hammer.

(1) calibration the sensor sensitivity.

(2) Trigger function on,

(3) AD over-cancel function, double hammer cancel function on,

(4) Press the FRF button to display the frequency response function.

(5) Press the SECOND button, then the PHASE button to switch to a two-screen display with the upper phase and lower MAG.

(6) Display the Bode plot.

You can also use the "impact test" option in EASY OPERATION.

Related instruction manual chapters

4.3.6 Frequency Response Function
5.3 calibration Function
3.7 Trigger Function
3.8 Double Hammer Cancel Function
3.2 AD Over-Cancellation Function
6.1. (3) Method using autoscaling
4.3.6. (3) Bode plot

Last updated: 2006-01-16