E-Mail: overseas@onosokki.co.jp

Select your region & language

Global

Region

CF-5200 Frequently Asked Questions

Question number FAQ-0584

Tracking analysis revealed that the DC component was removed. (DC coupling was selected.)

During constant ratio tracking
Because a tracking anti-aliasing filter (rotation speed tracking low-pass filter) is used, the internal input is AC coupled.

When using constant-width tracking
Since you specify a frequency range instead of an order setting (which results in a fixed anti-aliasing filter), tracking analysis is possible even for DC components.

Last updated: 2006-09-16