The procedure for tracking measurement is
The procedure for tracking constant ratio times is as follows
Please refer to the operation manual for the table of contents page (PDF).
Input the rotary pulse signal.
(3.3.6 Settings for External Input Signals)
(2) Check if the rotation speed is displayed correctly with TRACK−CONTROL−Rev.Disp ON.
Check if the degree is displayed on TRACK−CONTROL−ODER ANARYSIS.
Execution of Rotation Count Ratio Analysis
(3) Set the TRACK−SETUP MENU−TRACK DISP menu.
Set the menu for TRACK−SETUP MENU−ORDER SET.
Set the menu for TRACK CONDITION in the TRACK SETUP MENU.
(Execution of fixed ratio count and fixed width count tracking analysis)
(4) Turn TRACK−CONTROL−TRACK ON to on
Turn on ORDER ANALYSIS (constant ratio analysis)
or / turn off ORDER ANALYSIS (fixed-width analysis)
(Execution of fixed ratio count and fixed width count tracking analysis)
(5) ANALYSIS - AVERAGE - PER SP - EXP / Average Count Setting
(2.2.4 Setting of Averaging)
(6) Turn on the START button on the panel. (Measurement begins)
Refer to the 'Fixed Ratio and Fixed Width Tracking Analysis' (PDF).
Chapter of the relevant instruction manual
Execution of 3.1.2 rotation frequency ratio analysis
Execution of 3.3.2 constant ratio and constant width frequency tracking analysis
Execution of 3.3.2 constant ratio and constant width frequency tracking analysis
Setting of 2.2.4 averaging