CF-5200 Frequently Asked Questions
Question number FAQ-0573
The deceleration measurement is performed from 600 rpm to 0 rpm, but the measurement ends when the rotation display drops below 550 rpm.
As a cause
(1) The rotation pulse signal is not being captured properly.
Please ensure that the current rotation display is shown accurately.
(2) Slow down the change in rotation.
At low speeds, the sampling frequency is low, which means data sampling takes a long time, and therefore the FFT process also takes a long time.
Furthermore, measuring the rotational speed takes time, delaying the determination of whether the rotational speed required for data collection has been reached.
Since the minimum measurable rotational speed is 20 r/min, it is possible that the rotational speed may fall below this minimum during the sample collection process.
Related instruction manual chapters
3.6 Setting the External Input Signal
Last updated: 2006-09-16