CF-5200 Frequently Asked Questions
Question number FAQ-0567
Even when the upper limit for tracking data measurement is set to 10,000 r/min, the data is being collected at around 10,300 r/min.
For example, when starting data acquisition at 9999 r/min and completing 2048 points, if the rotation speed is 10300 r/min, the data acquisition is considered complete. The rotation speed will be the average rotation speed during that time.
Data is collected starting at 9999 r/min. Even if the rate drops below 9999 r/min during the data sample, data collection continues, but the data will not be used as tracking data unless it reaches 10000 r/min or higher.
Last updated: 2006-09-16