CF-5200 Frequently Asked Questions

Question number FAQ-0560

How to view the waveform of FD time data filtered from 500 to 1000 Hz?

(1) One method is to perform an FFT analysis on the FD data and use the bandwidth-limited IFFT (inverse Fourier transform) function.

(2) Another method is to combine two PS701 filter units and use them as a bandpass filter.

Related instruction manual chapters

7.2 Analysis of Memory Data
5.6 Inverse Fourier Transform

Last updated: 2000-04-10