CF-5200 Frequently Asked Questions
Question number FAQ-0506
The method for playing back uncalibrated noise data from FD and performing calibration is
(1) Set the sound level meter to the range when measured, re-calibration CF with the CAL signal, and read the FD data.
Use the overall value of the CAL signal as the correction value.
(Example) sound level meter 70dB range, CAL 64dB
CF Overalls = -9 dB
Add the correction value = 9 + 64 = 73 to the analysis data value (dB).