Question number FAQ-0462

Recording vibration and calibration signals (80Hz 1V) on the recorder. I want to perform real-time 1/1 OCT analysis, but 1) what is the calibration method? 2) what are the criteria for selecting the time constant?

Use the calibration function.

Set the Y-axis to LIN, perform spectral analysis on the CAL signal, and key in the overall value in m/s 2. (Function to set any search point value to any EU value.)

For max analysis, set the time constant to 10 ms.

For vibration level analysis, select a time constant of 630ms and a vibration filter.

Chapter of the relevant instruction manual

5.3.2 Set any voltage to 1 EU
5.3.3 Set any search point value to any EU value
5.8.4 A characteristic correction
Real-time analysis manual