CF-5200 Frequently Asked Questions

Question number FAQ-0461

I want to calibration the Y-axis by inputting a signal from a calibration with a frequency of 9.8 m/s² (= 1G) and displaying the single-amplitude value.

If the output signal of a 9.8 m/ calibration is displayed as an RMS value, the peak value will be √2 times the RMS value.

Please also change the CF display from RMS value after EU calibration to single amplitude value display.
It is convenient to perform calibration operation by spectrally analyzing the calibration signal and using the overall value to "set an arbitrary search point value to an arbitrary EU value."

Related instruction manual chapters

4.3.1. (1) Combinations based on the Y axis
5.3.2 Let any voltage be 1 EU.
5.3.3 Convert any search point value to any EU value