E-Mail: overseas@onosokki.co.jp

Select your region & language

Global

Region

CF-5200 Frequently Asked Questions

Question number FAQ-0434

We applied an EXP window and calculated the attenuation rate using curve fitting, but has the window been corrected?

The curve-fit attenuation rate is the value with the EXP window applied.
The window correction calculation formula is the same as that for the CF350 Hilbert transform. (Refer to the CF350/360 instruction manual.)

Related instruction manual chapters

Calculation of logarithmic decay rate using Hilbert transform in the CF350 instruction manual.

Last updated: 2006-01-16