CF-5200 Frequently Asked Questions
Question number FAQ-0377
To double-integrate the time waveform of acceleration G and calibration it to displacement (μm),
(1) In the EU calibration function sensitivity settings, set the voltage per 1m/s² and set the unit to "m".
(2) It displays displacement using the time axis integration function.
(3) In the old version, the speed/displacement direct reading function "G→mm" can be used.
In the new version, this function has been removed to comply with the revised measurement law.
Chapter of the relevant instruction manual
Set 5.3.11V to any EU
5.5.3 Speed/Displacement Direct Reading Function
5.5.2. (2) Frequency Axis Integration
Last updated date: 2006-01-16